6.1.2 Time control
The following settings can be made:
6.1.3 Configuring the Analog outputs (option)
1. Switch on instrument.
2. Select
“Remote setup” with Enter.
3. Select
“Analog outputs” with Enter.
4. Select desired value with cursor keys
(left/right).
5. Confirm with Enter.
6. Input 4 mA value via numerical keys.
7. Confirm with Enter.
8. Input 20 mA value via numerical keys.
9. Confirm with Enter.
Parameter
Meaning
Adjustment change
Interval time
Total time per cycle
Entry after activation
with Enter: 10 to 2550
minutes
Measuring
time
Time during which the measured
values are available. At the end of
the measuring time, a data record
is saved under 0x2000 and
following.
Entry after activation
with Enter: 5 to 60
minutes
Automatic
Switching on the automatic mode.
Operation without time control is
not permitted.
ON / OFF after
activation with Enter
Gas analysis
Adjustments
Control
Remote setup
Invalid measurement results
Sensor drift phenomena lead to higher inaccuracies.
Use the device only with time control!
CAUTION
31 Stat:
2 T.Gas 20.0 400.0 °C
12 CO2 0.00 20.00 %
4 O2 0.00 5.00 %
5 CO 0 500 ppm
6 NO 0.0 100.0 ppm
7 NO2 0.0 10.0 ppm
28 NOx 0.0 100.0 ppm