INMANUUIN121 M. USO VI340-HC118G2 ING USA V.2016.10.06.DOCX 5
3.3.2. PROCAM HD Camera
1
White Balance Button.
2
Anti-Moiré Button.
3
Window Button.
4
White Balance LED.
5
Anti-Moiré LED.
6
Window LED.
4. INSTALLATION
Connection and revision of this Equipment must be done by technical personnel with electrical installation
knowledge
The PROCAM HD includes the following components. Check and identify them before proceeding to the installation.
Console
Videocamera HD
Main power cable
Use video monitors certified under IEC60601-1 ed 3.0 or later
Use the endoscope in accordance with IEC60601-2-18
1.
Unpack the device with care.
2.
Place the PROCAM HD console on a solid surface.
3.
If the building structure has equipotential grounding, connect the equipment to ground through the equipotential ground terminal located
on the rear panel of the unit.
4.
Connect the connector of the video camera ECLERIS to the connector in the front panel.
5.
Plug the video output connector on the rear panel of the PROCAM HD console to the video input of the video monitor.
Check if the image device (videomonitor / video printer / LCD TV etc…) supports the camera resolutions
6.
Connect the optical device to the camera head with the C-Mount screw (Endoscope, Microscope, Colposcope, etc)