HYDRA-DS NO
3
-N
Page 16
2.3
M
OUNTING
2.3.1
HYDRA
S
ENSOR
M
OUNTING
Install the sensor where the measured sample is representative of the entire process. Although the sensor will function
in a quiescent sample, flow improves the measurement. The recommended minimum flow is 0.1 m/sec.
Securely mount the HYDRA-
DS sensor with the measuring end at least 6” away from the tank wall and
bottom. Ensure
that the sensor is immersed at least 6” at all times.
Use care when servicing the sensor to ensure that the sensor does not hit the tank wall or bottom, which could break
the sensing electrodes.
2.3.2
T80
A
NALYZER
M
OUNTING
Mount the T80 in a location where there is easy access to the analyzer and sensor. Install the system in an area where
vibrations, electromagnetic, and radio frequency interference are minimized or absent. So not mount in direct sunlight
or areas of extreme heat (temperature > 120°F). The IP65 T80 is suitable for outdoor use but it is best to mount it with
a protective cover or sunshield to prevent discoloring over the years.
Universal Mounting Brackets
Rail Mounting