The Uninterruptible Power Supply (UPS) compensates for voltage
fluctuations and provides backup power for the computer during
power failures for a maximum of five minutes.

The UPS is unable to power the acquisition system during power
failure. Any measurements that are running on the computer are
stopped.

UPS front panel

The UPS front panel provides buttons and a control panel. Useful
information is displayed about the UPS, load status, events,
measurements, and settings.

3

3.1

Description

3

SOMATOM | Version 1

13

Print No. HC-C2-X05.610.64.01.02

Summary of Contents for SOMATOM

Page 1: ...www siemens com healthcare SOMATOM Instructions for Use Uninterruptable power supply UPS Version 1...

Page 2: ......

Page 3: ...SOMATOM Instructions for Use Uninterruptable power supply UPS Version 1...

Page 4: ...and keys for example the Save button Is used to emphasize particularly important sections of the text Is used for on screen output of the system including code related elements or commands Is used to...

Page 5: ...death or serious injury WARNING consists of the following elements Information about the nature of a hazardous situation Consequences of not avoiding a hazardous situation Methods of avoiding a hazard...

Page 6: ...Legend 6 Uninterruptable power supply UPS Instructions for Use Print No HC C2 X05 610 64 01 02...

Page 7: ...1 UPS front panel 13 3 1 1 Buttons 14 3 1 2 Control panel 16 4 Operation 19 4 1 Normal mode 19 4 2 Battery mode 19 4 3 Shutting down the UPS 20 5 Technical specifications 21 5 1 Dimensions and weight...

Page 8: ...Table of contents 8 Uninterruptable power supply UPS Instructions for Use Print No HC C2 X05 610 64 01 02...

Page 9: ...Definition AS SOMATOM Definition Edge SOMATOM Definition Flash SOMATOM Emotion SOMATOM Emotion Duo SOMATOM Emotion 6 SOMATOM Emotion 16 SOMATOM Sensation SOMATOM Sensation 10 SOMATOM Sensation 16 SOMA...

Page 10: ...s document is valid only in conjunction with the Instructions for Use of the corresponding SOMATOM CT system 1 2 1 Introduction 10 Uninterruptable power supply UPS Instructions for Use Print No HC C2...

Page 11: ...e instructions contained therein Read the safety instructions in the CT system documentation in addition to the following safety information For more information see the SOMATOM Instructions for Use S...

Page 12: ...rmation about disassembly of the unit CAUTION Improper disposal of the system or parts of the system Pollution of the environment System components hazardous to persons or the environment must be disp...

Page 13: ...nutes The UPS is unable to power the acquisition system during power failure Any measurements that are running on the computer are stopped UPS front panel The UPS front panel provides buttons and a co...

Page 14: ...tton 5 Enter button 6 Down button 7 Up button 8 Escape button Buttons The following table displays the buttons and their functions on the LCD Button Name Function ESC Exits the menu 3 1 1 3 Descriptio...

Page 15: ...s the UPS on or off The following table displays the front panel items and status descriptions Item Color Status Description green on The UPS is operating normally yellow on The UPS is in battery mode...

Page 16: ...Press any button to restore the screen The following table describes the status information provided by the UPS Icon Operation status Possible cause Action Standby mode The UPS is off waiting for the...

Page 17: ...evice The equip ment is still protected On Battery 1 beep every 10 seconds A utility failure has occurred and the UPS is in bat tery mode The UPS is powering the equipment with battery power Pre pare...

Page 18: ...3 Description 18 Uninterruptable power supply UPS Instructions for Use Print No HC C2 X05 610 64 01 02...

Page 19: ...vided by the battery The audio alarm beeps every ten seconds while the equipment connected to the UPS is supplied by the battery The power indicator and the battery indicator illuminate continuously T...

Page 20: ...S Press the Power button on the front panel for three seconds The UPS starts to beep and shows the status UPS shutting OFF The UPS then transfers to standby mode and the power indicator turns off 4 3...

Page 21: ...150 mm 5 9 inch 3 Height 233 mm 9 2 inch 3 Length 445 mm 17 5 inch 3 Weight 15 95 kg 35 20 lb 3 Storage If the UPS must be stored prior to installation storage needs to be in a dry place The permitte...

Page 22: ...5 Technical specifications 22 Uninterruptable power supply UPS Instructions for Use Print No HC C2 X05 610 64 01 02...

Page 23: ...r s identification label 2 Warning label PROP 65 WARNING Battery posts terminals and related accessories contain lead and lead compounds chemicals known to the state of California to cause cancer and...

Page 24: ...for installation in a controlled environment For maintenance and disposal instructions of the battery see user s manual CAUTION Lead Acid battery inside the enclosure Risk of electrical shock due to h...

Page 25: ...ble systems 9 control panel 16 F front panel 13 L location of labels 23 N normal mode 19 O operation 19 S safety information 11 shutting down UPS 20 T technical specifications 21 Index SOMATOM Version...

Page 26: ...This page has been intentionally left blank 26 Uninterruptable power supply UPS Instructions for Use Print No HC C2 X05 610 64 01 02...

Page 27: ......

Page 28: ...arters Siemens Healthcare GmbH Henkestr 127 91052 Erlangen Germany Phone 49 9131 84 0 siemens com healthcare Legal Manufacturer Siemens Healthcare GmbH Henkestr 127 91052 Erlangen Germany Print No HC...

Reviews: