XS series operating manual
XS user manual v0.55
- 146 -
Operation of the XS 4-wire option in XS hipot function
Reminder:
The purpose of this option is to detect the presence of the test piece and not to improve
measurement accuracy
(like the 4-wires in ground continuity)
The principle implemented is based on the use of an additional HV cable and a
2nd GND return cable for voltage re-reading. Detection of the absence of a test piece corresponds to a
nil voltage re-reading.
Option operation phase
Warning
The option is only active during the test hold phase.
The user wishing to save time in the event of the non-presence of a test piece can stop the test at the
beginning of the hold phase, as soon as the instrument status switches to connection fault.
Fall
Rise
Time
Voltage generated
Hold
Option active