EASYHOME EBP-020 User Manual Download Page 1

EBP-020

Blood Pressure Monitor 

User Manual

Summary of Contents for EBP-020

Page 1: ...EBP 020 Blood Pressure Monitor User Manual...

Page 2: ...by this equipment are equivalent to those measured with the auscultatory method The auscultatory method is auscultatory method where a trained healthcare provider uses a blood pressure meter and liste...

Page 3: ...Device 4 5 Measurement 7 6 About Blood Pressure 11 7 Contraindications Precautions Warnings and Prompt Instructions 13 8 Q A 17 9 Troubleshooting 19 10 Cleaning and Disinfection 21 11 Storage 23 12 Pr...

Page 4: ...pulse rate of an adult via the non invasive oscillometric method Oscillometric Method is measured by establishing the relationship between systolic blood pressure diastolic blood pressure mean blood p...

Page 5: ...ments When you look at historical records the first value that appears is the average of the latest three measurements 2 1 2 KPa mmHg M AVG User icon Date time display Unit of Measurement Blood Pressu...

Page 6: ...method shown in the figure above If the low battery symbol appears on the screen after the product is turned on the measurement cannot be performed and the batter ies must be replaced b Place 4 AAA d...

Page 7: ...void battery corrosion and danger to device Optional Not Included The product s power can be supplied through a DC 5V external power supply through the USB port 4 Cuff The arm circumference rang of th...

Page 8: ...ill display xx xx and flash the month d Press the Memory key to increase the value between 1 and 12 Press the User key for confirmation and enter the date setting mode After entering the date setting...

Page 9: ...isappear to restore the factory settings But the historical data measured will not be lost If this happens please follow the above steps to reset year month day time In this mode the SYS area will dis...

Page 10: ...carried out in a quiet environment Avoid eating smoking or any form of heavy physical labor as it will affect the measurement result Sit on a chair to have a rest for about 10min in a quiet environmen...

Page 11: ...re in a comfortable and relaxed state before measurement Do not move or exercise the muscles of your arm during the measure ment If necessary a cushion can be used to support your arm Be sure the cuff...

Page 12: ...t Please note that the rubber tube on the cuff cannot be folded or bent Note If your left arm cannot be used for measurement please use your right arm But in any case all measurements must be made on...

Page 13: ...en inflated to a stable pressure it will stop inflating The pressure in the cuff will decrease gradually and be displayed on the screen When the air pressure is insufficient the machine will automatic...

Page 14: ...us measurements starting with your very first measurement Continue to press the Memory Key to scroll through all of your measurements Press the User key to switch to look through the other user s data...

Page 15: ...of two forces The first force systolic pressure occurs as blood pumps out of the heart and into the arteries that are part of the circulatory system The second force diastolic pressure is created as t...

Page 16: ...old should be accompanied by an adult when taking measure ments This device has not been tested for use by pregnant women Hypotension low blood pressure has not been defined It is general ly considere...

Page 17: ...f following the user manual and pay special attention to the use of the hose to avoid twisting or twining due to excessive length Do not use the cuff for a long period of time to avoid a possible alle...

Page 18: ...his model of blood pressure monitor is suitable for the arm circumference range of 22 42cm If your arm is beyond this range you might not obtain an accurate measurement This model of blood pressure mo...

Page 19: ...peratures exceeding room temperature up to 131 wait 1 hour with the device in a room temperature environment before use Do not pull the cuff forcefully or use the cuff as a rag Do not be contaminate w...

Page 20: ...leeding or severe arteriosclerosis or elderly with poor peripher al circulation particularly patients with peripheral vascular lesion caused by complicated severe diabetes In case of rescue of serious...

Page 21: ...red 1 Because the blood pressure is constantly changing it is difficult to obtain the correct blood pressure value by measuring only once Please take the average measurement of the last three times fo...

Page 22: ...shooting LCD display information Reason Meaning Err is displayed in the SYS area 1 is displayed in the DIA area The leakage is too fast or the pulse signal is too weak Please check the cuff tie it up...

Page 23: ...batteries reversed Replace the batteries with new ones or plug the machine into a Micro USB for power supply Install in accordance with the positive and negative poles of the batteries No pressure Is...

Page 24: ...quid to prevent liquid from flowing into the machine 10 Cleaning and Disinfection 21 The cuff leaks Is the cuff wrapped too loose Please tighten the cuff The inflatable air cuff bag of the cuff is bro...

Page 25: ...od pressure monitor before and after use each time Each disinfection shall be completed within 1min The number of repeated disinfection each time shall not exceed 2 times Cleaning and disinfection sha...

Page 26: ...isten the cuff Diluent Volatile oil Gasoline Do not store the product at a place which has direct sunlight high temperature humidity dust or corrosive gas If the Blood pressure monitor is not in use f...

Page 27: ...020 Pulse Memory capacity 99 groups of measurement data for each of two users Power supply 4 AAA dry batteries or DC 5V USB external power supply Dimension 120mm length 90 width 43mm height Accessori...

Page 28: ...humidity range it will not work properly Strong impact direct collision exposure to the sunlight or rain should be avoided during transportation The packaged blood pressure monitor should be stored i...

Page 29: ...electromagnetic environment specified below The customer or the user of the Model EBP 020 Blood pressure monitor should assure that it is used in such an environment The Model EBP 020 Blood pressure...

Page 30: ...150 kHz to 80 MHZ outside ISM bandsa ower frequency magnetic fields should be at levels characteristic of a typical location in a typical commercial or hospital environment Power frequency 50 60 Hz ma...

Page 31: ...strial scientific and medical bands between 0 15 MHz and 80 MHz are 6 765 MHz to 6 795 MHz 13 553 MHz to 13 567 MHz 26 957 MHz to 27 283MHz and 40 66 MHz to 40 70 MHz The amateur radio bands between 0...

Page 32: ...m output power of the communications equipment For transmitters rated at a maximum output power not listed above the recommended separation distance d in meters m can be estimated using the equation a...

Page 33: ...d for Easy Healthcare Corporation 360 Shore Dr Unit B Burr Ridge Illinois 60527 UNITED STATES Questions or comments Please call toll free 1 855 822 6999 M F 9 a m 5 p m CST E mail service healthcare m...