![East Tester ET3260 User Manual Download Page 51](http://html1.mh-extra.com/html/east-tester/et3260/et3260_user-manual_3438427051.webp)
Users manual
ET3260 and ET3260A
6 1/2 Digit Multimeter
51
6.4.2 Low-frequency additional error
Frequency
Gate time (resolution)
1 second (6 1/2)
0.1 second (6 1/2)
0.01 second (6 1/2)
3Hz-5Hz
0
0.12
0.12
5Hz-10Hz
0
0.17
0.17
10Hz-40Hz
0
0.20
0.20
40Hz-100Hz
0
0.06
0.21
100Hz-300Hz
0
0.03
0.21
300Hz-1kHz
0
0.01
0.07
>1kHz
0
0
0.02
6.5 Capacitance
Accuracy is expressed by ± (% of measurement value+% of measurement range) [1]
Function
Measurement
range
Test current
Resolution
Display speed Annual accuracy
% of the reading
+
%
of
measurement
range
Temperature
coefficient % of the
r % of
measurement range
Capacitance
measurement
1nF
10uA
1pF
≥2sps
2+2.5
0.05+0.05
10nF
10uA
10pF
≥2sps
1+0.3
0.05+0.01
100nF
100uA
100pF
≥2sps
1+0.3
0.01+0.01
1uF
1mA
1nF
≥2sps
1+0.3
0.01+0.01
10uF
1mA
10nF
≥2sps
1+0.3
0.01+0.01
100uF
1mA
0.1uF
≥2sps
1+0.3
0.01+0.01
1000uF
10mA
1uF
≥2sps
1+0.3
0.01+0.01
10mF
10mA
0.01mF
≥2sps
1+0.3
0.01+0.01
100mF
10mA
0.1mF
≥2sps
3+0.2
0.05+0.02
[1] Pre-heat for 60 minutes, and use
Relative
operation function.
6.6 Temperature
Accuracy is expressed by ±
℃
[1]
Function
Type
Test current
Best range
Display speed
Annual
accuracy
Temperature coefficient
Temperature
test
Pt100
1mA
-200 to 660℃
≥2sps
0.16°C
0.01°C
Pt1000
100uA
-200 to 660℃
≥2sps
0.16°C
0.01°C
[1] Pre-heat for 60 minutes, technical indicator does not include probe accuracy, and it has to add
probe accuracy during calculation.
[2] Indicator means that sensor uses four-wire resistance measurement or two-wire resistance
measurement using
Relative
operation.