13
6 Advanced Application
6.1 Setting 0 % and 100 % output
parameters
As for stepping operation and percentage display, 0% and 100% should be set before
using. Some gear values have been set when delivered from the factory and set values
are illustrated below:
Output function
0% value
100% value
DC mV
0.00 mV
100.00 mV
Thermocouple J type
0.0
℃
1000.0
℃
Thermocouple K type
0.0
℃
1000.0
℃
Thermocouple T type
0.0
℃
400.0
℃
Thermocouple E type
0.0
℃
800.0
℃
Thermocouple R type
0
℃
1500
℃
Thermocouple S type
0
℃
1500
℃
Thermocouple B model
600
℃
1800
℃
Thermocouple N model
0.0
℃
1000.0
℃
The default set values may not meet your requirements, so which can be reset. Then
you can display the percentage with stepping or slope output as well as display the
percentage. Adjust the output value with keys
, press
or
for a
long time until the buzzer works to define the new 0% or 100% values. The new
defined span value is saved in the calibrator storage automatically and remains
effective after restart. The flowing operations are available with the setting:
Press
or
for manual steeping (increase or decrease) output based