background image

Catalogue

1.

Fault code list

------------------------------------------------------------ 3-4

2.

List of tools to use

------------------------------------------------------------ 5

3.

Cloak replacement -

------------------------------------------------------------6

4.

Replacement of flannelette in backrest

----------------------------------------- 7

5.

Movement disassembly

-1 ----------------------------------------------------- 8

6.

Movement replacement

-2 --------------------------------------------------- -9

7.

Movement replacement

-3 --------------------------------------------------- 10

8.

Remove the back cover

--------------------------------------------------------11

9.

Replacement of upper cover of central box

--------------------------------------12

10.

Remove side plate

----------------------------------------------------------- 13

11.

calves replacement-

---------------------------------------------------------- 14

12.

Movement backrest bus disassembly

-1 ------------------------------------- 15

13.

Movement backrest bus disassembly

-2------------------------------------- 16

14.

Movement backrest bus disassembly

-3---------------------------------------- 17

15.

Gas pump replacement

------------------------------------------------------- 18

16.

centre plate

transformer

Fuse replacement

---------------------------------- 19

2

Summary of Contents for EI-8602A-TIT

Page 1: ...Xiamen Healthcare Electronic Co Ltd Maintenance Service Manual Model EI 8602A TIT Voltage Specification 110 120V By the Easepal s Customer Service Dept 1...

Page 2: ...cement 2 9 7 Movement replacement 3 10 8 Remove the back cover 11 9 Replacement of upper cover of central box 12 10 Remove side plate 13 11 calves replacement 14 12 Movement backrest bus disassembly 1...

Page 3: ...3 Replace the center board 3 The movement plate does not communicate with the center board for more than 4s Bad contact or fracture of inserting wire 1 Check or replace the plug in cable 2 Replacemen...

Page 4: ...acement of sensors 3 Replacement of movement board 1 Detection when function is turned on 2 Shut down automatically in the event of failure time And not allowed to boot 54 The position of the Core kne...

Page 5: ...Tools List Diagonal Cutter Electric Screw Driver Electric Screw Driver Cable Ties Cross Batch 2 6 35 65 Slot Type Screwdriver 2 6 35 100 Screwdriver 5...

Page 6: ...Pass through the zipper head with a small Cable Ties and pull it apart figure 1 as shown in figure 2 after opening at 7th position 2 Put on the new cloak and zip up 7th position zippers as shown in fi...

Page 7: ...the first step 2 Open 4 zippers of flannelette according to figure 1 3 Remove the two screws behind the back and the two screws under the pedestal As shown in figure 2 4 Replace the new flannelette w...

Page 8: ...hole to move the machine core away from the guideway to remove the baffle area If the machine core does not block the baffle area of the guideway jump directly to the next step 3 Remove the baffle fr...

Page 9: ...4 The walking gear comes out first and the front gear comes out again 8...

Page 10: ...wn in figure 3 with Diagonal Cutter 6 Remove the earth wire of the machine core 7 Pull out the trachea and terminal 8 Connect the trachea to the terminal with a new machine core 9 Lock the earth wire...

Page 11: ...2 The screwdriver removes the machine core away from the baffle area of the left and right guideway through the repair hole 13 install the Remove baffle of the left and right guideway on the guideway...

Page 12: ...the power supply 2 Pull out the cloth pin on the back cover vertically as shown in figure 1 and remove the four screws in figure 2 3 Press down pull the back cover backward and pull out a certain ang...

Page 13: ...5 cloth pins in turn perpendicular to the direction of the upper cover as shown in figure 1 3 Remove the screws as shown in figures 2 and 3 4 Hold the inner side of the top cover of the center box on...

Page 14: ...ewdriver as shown in figure 4 6 Connect the power supply run the massage chair to the back state remove the screw at the front end of the side plate from below the calves as shown in figure 5 and then...

Page 15: ...of calves 1 Unzipper as shown in figure 1 2 Break the DK Card bolt open with your fingers or pliers and take it up as shown in figure 2 3 Remove the trachea and power cord connector as shown in figure...

Page 16: ...or removing the cloak and velvet through P7 step 1 3 2 Press P11 to remove the rear cover cut off the tie tape shown in the figure below pull out the terminal shown in the figure below and remove the...

Page 17: ...wire 3 places The earth wire of the nine core connecting wire is locked to the rear cover of the movement One end of the spring is pierced into the buckle and fixed with the reserved band Pull spring...

Page 18: ...line is connected to the 32 core wire 17 8 10 9 Movement backrest bus disassembly 3 Nine core connecting wire and 32 core wire earth wire with lock Straighten the line tie the belt and fix it 7 Cloth...

Page 19: ...pump replacement 1 Remove 5 cloth sleeve pins on the top cover of the central box 2 Remove 3 screws 3 Remove 4 screws of the air pump and cut off the bandage so that you can take it out and replace it...

Page 20: ...plug the terminal Remove the fixing screw from the transformer Remove transformer 2 Center plate replacement Unplug the terminal from the center board Remove 6 screws That is the center board can be r...