After Electrical Power is restored
1. Make sure that the InSight Device is attached to the Cradle.
2. Make sure the Cradle is connected to the power socket.
3. The system automatically starts operating
4. Ensure that the InSight Device's details are appearing correctly upon the RCD Unit Screen.
2.6
Clinical Alert
Alerts are used to notify personnel of situations which may require attention. The EarlySense
InSight
Device alert’s the user that an Alert is active by
flashing a LED light, or/and by making an
audible sound. The color of the LED light indicates the type of alert (alert types are specified in
the RCD Unit's User Guide). To understand the nature of an Active Alert, the user must view the
RCD Unit's User Guide.
An active Alert can be acknowledged either by pressing on the InSight Device Front Panel, or
via the RCD Unit. The InSight System can be also preset to acknowledge alerts via the InSight
Device only.
Vital Signs Alerts are generated by the system if heart rate or respiratory rate averages exceed
predefined thresholds for the amount of time listed below:
Low Heart Rate
High Heart Rate
Time to alert for change in HR
90 seconds
90 seconds
Low Respiratory Rate
High Respiratory Rate
Time to alert for change in RR
180 seconds
180 seconds
The InSight System can also be configured to provide a "Multi Parameter Alert". This Alert is
generated if both HR and RR exceeds predefined thresholds, as set by the EarlySense
technical support representative, according to the Healthcare practitioner's instructions.
WARNING
The Audible and Visual Alerts in the EarlySense InSight System are not intended
for monitoring high risk situations where ECG monitoring is required. The most
reliable method of patient monitoring combines close personal surveillance
with correct operation of the monitoring equipment.
2.7
Technical alerts
Technical Alerts are managed differently than other alerts. Technical Alerts indicate that there is
a technical problem. Please refer to the Remote Control and Display (RCD) Unit User Manual to
understand the nature of the alert.
2.8
Maintenance and cleaning
The EarlySense InSight System is designed to provide trouble free operation. The system should
be kept clean and dry, and the user should verify that there is no physical damage to the InSight
Device and to the Bed/Chair Sensing Units.