![e-LinkCare UBREATH PRO PF680 Manual Download Page 47](http://html1.mh-extra.com/html/e-linkcare/ubreath-pro-pf680/ubreath-pro-pf680_manual_3099658047.webp)
45
Calibration
When the
UBREATH® PRO
Spirometer System is in use, the users should always
verify its accuracy by frequently checking its performing maintenance procedures
at regular intervals (not required, but optional).
The following picture shows the “Calibration" screen.
You may need a syringe and adaptor separately. Contact your local distributor for
more information.
The
UBREATH® PRO
Spirometer System calibration simulates inspirations and
expirations for same flow levels each to ensure the product performance.
Follow the below procedure to complete the calibration:
1.
Tap on "3L” button, you will have three options to choose from different
volume of syringes, select the right syringe that you are using. Press “Back”
button to go back to previous menu.
2.
Connect the syringe and the adaptor to the new
UBREATH® PRO
Disposable Flow Transducer, avoid any cross infection.
3.
Press “Start” button to initiate a quality control, a following screen will pop
up to guide you through the process.