Installation

Pressure Connection

Use a Ø7.5 mm drill for installing the pressure connection nipples into the duct.

Installation Examples

1) Measuring range    2) Display backlight    3) These and further settings can be changed with PCS10 via USB configuration adapter (HA011066) while DIP 

switch S9 = 1. Only available with version with 10 DIP switches.

S1

S2

MR

1)

S3

S4

Time

S5

S6

Unit

S7 DPB

2)

S8

Output

S9

Setting

3)

0

0

100 %

0

0

50 ms

0

0

Pa

0

on

0

0 - 10 / 4 - 20 mA

0

 

DIP switch

1

0

75 %

1

0

500 ms

1

0

mbar

1

off

1

0 - 5 / 0 - 20 mA

1

PCS10

0

1

50 %

0

1

2 s

0

1

inch WC

1

1

25 %

1

1

4 s

1

1

mm H

2

O

Red LED 

Flashing (1s interval)

= The measured data is out of the 

selected range (overload or reversed 

pressure connection)

One flash (2s) 

= Indicates the failure of the attempt to 

adjust zero point or span point, or to 

return to factory adjustment

Green LED 

Flashing (1s interval)

= EE610 operates normally, the 

measured data is within the selected 

measuring range

One flash (2s) 

= Confirms adjustment or return to 

factory settings

Off

=

No power supply or 

electronics failure

Fast flashing

 

(0.2 s interval)

= Auto-zero is executed, first time 10 s 

after  start / reset

Connection Diagram

V+

GND

V+

A (=D+)

RS485

B (=D-)

GND

AV

1
2
3

1
2
3
4

+

-

~

~

+

-

~

~

V / mA

Power supply

24 V AC ±20 %

15...35 V DC

Power supply

15 - 35 V DC

24 V AC ±20 % 

Important note:

  