User Manual EE360
Moisture in Oil Sensor | 18
5 Optional Modules
5.1 Alarm Module (Option AM2)
The module offers two freely configurable relay outputs for alarm or control purposes. Various operation modes
are available including switch hysteresis, switch window and error indication. The error modes can be configured
independently from each other. The measurands at the outputs as well as switching points, hysteresis and the
normal state (standard / inverted) can be set via PCS10 Product Configuration Software or using the push-
buttons (see chapter 10.3 Optional Menu, Fig. 40 Alarm output).
Switch Hysteresis Mode
The switching behavior is determined by entering a switching point and an associated hysteresis value.
t
closed
open
closed
open
Normal state: Standard
Normal state: Standard
Terminal 1-2 / 4-5
Terminal 2-3 / 5-6
Switching point
Hysteresis
RH [%]
Fig. 9
Example hysteresis mode on both alarm outputs using
normal state standard
Additionally, detected errors are signaled at the alarm output.
t
closed
open
closed
open
Error occurs
No error
RH [%]
Switching point
Hysteresis
Terminal 2-3
Terminal 5-6
Normal state: Standard
Normal state: Inverted
Fig. 10 Example of hysteresis mode with error occurring.