32
EE33 Humidity / Temperature Sensor for High Humidity and Chemical Applications
10.3 Fuse Replacement for Integrated Power Supply (Option AM3)
If the green LED on the PCB is not flashing with the supply
voltage switched on, check the fuse and replace if required.
Fuse secondary:
250mA / T UL248-14
Nominal voltage:
250V
Replacement types:
Series: MSTU 250
Manufacturer: Schurter Order No.: 0034.7109
Series: 374
Manufacturer: Littelfuse Order No.: 374 0250.
10.4 Self Diagnosis and Error Messages
10.4.1 Self diagnosis via LEDs on the circuit board
LED D2 (green)
Flashing
Supply voltage applied / Microprocessor is active
LED D1 (red)
Constantly lit
Humidity sensor element damaged
Flashing
Dew (condensation) at the humidity sensor element
LED D3 (orange)
Constantly lit
Humidity sensor element will be heated (Automatic ReCovery)
Flashing
Sensor and threaded element soiled
10.4.2 Self diagnosis via display (optional):
Error 1
Humidity sensor element damaged
Error 2
Humidity sensor element moistened (condensation!)
Error 3
Temperature sensor element damaged
Error 4
Temperature sensor short-circuit
Error 5
Pt1000 probe element is damaged
Error 6
Pt1000 probe short-circuit
10.4.3 Further self diagnosis:
Error
Possible cause
Measures / Help
Display shows incorrect values
Error during re-adjustment of the transmitter
Reset to factory calibration and repeat the calibration routine
Filter soiled
Replace filter
Measuring cell contaminated
Automatic ReCovery (ARC)
Output configured incorrectly
Check output range and output signals in the configuration
Long response time
Filter soiled
Replace filter
Incorrect filter type
Filter type should match the application
Types T17, T28