29
User Manual EE300Ex Humidity / Temperature Sensor
7 Technical Data
7.1 EE300Ex-M1 Humidity and Temperature Sensor
Measurands
Relative humidity
Measuring range
0...100 % RH
Accuracy
1)
(including hysteresis, non-linearity and repeatability,
-15...40 °C
(5...104 °F)
≤90 % RH ± (1.3 + 0.3%*mv) % RH
traceable to international standards, administrated by NIST, PTB, BEV...)
-15...40 °C
(5...104 °F)
>90 % RH ± 2.3 % RH
-25...70 °C
(-13...158 °F)
± (1.4 + 1%*mv) % RH
mv = measured value
-40...180 °C
(-40...356 °F)
± (1.5 + 1.5%*mv) % RH
Temperature dependence electronics, typ.
0.03 % RH/°C
Response time t
90
< 30 s with stainless steel filter at 20 °C
(68 °F)
Temperature
Measuring range
Wall mount: -40...60 °C
(-40...140 °F)
Remote probe: -40...180 °C
(-40...356 °F)
For TIIS (Japan):
Model T1, T7, T10: -40...60 °C
(-40...140 °F)
Accuracy
1)
∆
T [°C]
T [°C]
-40
0
-10
-20
-30
10
50
40
30
20
60
100
90
80
70
140
130
120
110
150
170
160
180
∆
T [°C]
T [°C]
∆
T [°C]
T [°C]
Temperature dependence of electronics, typ.
0.005 °C/°C
Calculated parameters
Outputs
Freely selectable and scalable outputs
2 x 4-20 mA (2-wire) galvanically isolated
R
L
= (V
cc
-9V)/20mA
Output 1 must be connected!
General
Supply voltage
V
cc min
= (9+R
L
*0.02) V DC V
cc max
= 28 V DC
R
L
= load resistor
Current consumption
Max. 20 mA per channel
Protection class of housing
IP65 / NEMA 4
Cable gland
M16 for cable diameter 5 - 10 mm
(0.2" - 0.4")
M20 for cable diameter 10 - 14 mm
(0.4" - 0.6")
Electrical connection
Screw terminals max. 1.5 mm²
(AWG 16)
Working temperature range Probe
according measuring range
Electronics without display -40...60 °C
(-40...140 °F)
Electronics with display -20...60 °C
(-4...140 °F)
Storage temperature range Electronics and probe
-20...60 °C
(22...140 °F)
Electromagnetic compatibility
EN 61326-1 EN 61326-2-3
ICES-003 ClassB
Industrial Environment FCC Part15 ClassB
Material
Enclosure
stainless steel 1.4404
Probe cable
PTFE
Probe (without filter)
stainless steel 1.4301
1) The accuracy statement includes the uncertainty of the factory calibration with an enhancement factor k=2 (2-times standard deviation). The accuracy was calculated in
accordance with EA-4/02 and with regard to GUM (Guide to the Expression of Uncertainty in Measurement).
from
up to
Units
wall mount
remote probe
Dew point temperature
Td
-40
(
-40)
60
(
140)
100
(
212)
°C
(°F)
Frost point temperature
Tf
-40
(
-40)
60
(
140)
100
(
212)
°C
(°F)
Wet bulb temperature
Tw
0
(
32)
60
(140)
100
(
212)
°C
(°F)
Water vapour pressure
e
0
(
0)
200
(
3)
1 100
(
15)
mbar
(psi)
Mixing ratio
r
0
(
0)
425
(
2 900)
999
(
9 999)
g/kg
(gr/lb)
Absolute humidity
dv
0
(
0)
150
(
60)
700
(
300)
g/m³
(gr/ft³)
Specific enthalpy
h
0
(
0)
400
(
150 000)
2 800
(
999 999)
kJ/kg
(Btu/lb)
Water activity
aw
0
-
1
1
Water content
X
0
-
100 000
[ppm]
Summary of Contents for EE300Ex
Page 31: ...31 User Manual EE300Ex Humidity Temperature Sensor 8 ATEX Certificate...
Page 32: ...32 User Manual EE300Ex Humidity Temperature Sensor...
Page 33: ...33 User Manual EE300Ex Humidity Temperature Sensor...
Page 34: ...34 User Manual EE300Ex Humidity Temperature Sensor...
Page 35: ...35 User Manual EE300Ex Humidity Temperature Sensor 9 EU Declaration of Conformity...
Page 51: ...51 User Manual EE300Ex Humidity Temperature Sensor 13 Control Drawing M1_1309080...
Page 52: ...52 User Manual EE300Ex Humidity Temperature Sensor...
Page 53: ...53 User Manual EE300Ex Humidity Temperature Sensor...
Page 55: ...55 User Manual EE300Ex Humidity Temperature Sensor...