2.1.1  Sensing Probes

The EE1900 features a plastic (PPS) probe according to the order which can be employed over the 

entire T range according to the technical data. 

2.1.2  Filter Caps

The right choice of filter cap is very important for optimal performance under certain environment 

conditions. Please see details in the “Accessories” data sheet at www.epluse.com/EE1900 or contact 

the manufacturer’s representative for advice. 

2.2  Dimensions

All values are given in mm / inch.

Ø

 12 / 0.47

Probe length

(see ordering guide)

Cable length (see ordering guide)

Filter length

(see table below)

Ø 3.5 / 0.14

Ø 3.5 / 0.14

Fig. 1 

 Dimensions of EE1900

2.3  Installation

ESD handling regulations must be strictly observed. 

2.3.1  Electronics Board

Hold the electronics boards only by the edges. Do not touch the components or their contacting.  

2.3.2  Sensing Probe

Avoid touching the sensing head. The probe must be operated with the filter cap on at all times. 

A clogged filter cap causes longer response time. For replacing the filter cap and cleaning the sensing 

head please see the “Cleaning Instructions” at www.epluse.com/ee1900.

Do not touch the sensing elements!

Order code

Filter length

 

[

mm/inch

]

 F9

 39

 / 1.54

 F4

 33

 / 1.30

 no code

 33

 / 1.30

   F12

 33

 / 1.30

Summary of Contents for EE1900

Page 1: ...BA_EE1900 v1 1 Modification rights reserved User Manual EE1900 HUMIDITY MODULE FOR CLIMATE CHAMBERS...

Page 2: ...equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmf...

Page 3: ...al 5 2 1 1 Sensing Probes 6 2 1 2 Filter Caps 6 2 2 Dimensions 6 2 3 Installation 6 2 3 1 Electronics Board 6 2 3 2 Sensing Probe 6 3 Electrical Connection 7 4 Setup and Configuration 8 4 1 Selection...

Page 4: ...commissioning shall be performed by qualified personnel only 1 2 2 Mounting Start up and Operation The device has been produced under state of the art manufacturing conditions has been thoroughly test...

Page 5: ...E E Elektronik are developed and manufactured observing of all relevant requirements with respect to environment protection Please observe local regulations for the device disposal For disposal the i...

Page 6: ...47 Probe length see ordering guide Cable length see ordering guide Filter length see table below 3 5 0 14 3 5 0 14 Fig 1 Dimensions of EE1900 2 3 Installation ESD handling regulations must be strictly...

Page 7: ...tor If the probe is installed into a partition wall the probe s backend shall be isolated thermally 3 Electrical Connection Supply Fig 2 Power supply 15 35 V DC 17 29 V AC Connection supply and output...

Page 8: ...RH and error indication according NAMUR see chapter 4 3 is disabled After switching the output between voltage and current with the slide switch the scaling of the output can be set using the E E Pro...

Page 9: ...manually by pressing the Sensor Recovery button No 6 Fig 3 for min 3 seconds or automatically by applying a voltage of 5 30 V DC to the potential free input No 3 Heat IN Fig 3 for at least 3 seconds A...

Page 10: ...overy The ARC mode is indicated by the closed contact at the optocoupler output No 3 Status Fig 3 and by the red LED flashing very quickly The optocoupler output thus indicates wether the data on the...

Page 11: ...guration PC HA011017 EE1900 The adjustment shall be performed for the RH output against an appropriate RH reference See Calibration Kit User Guide at www epluse com ee1900 7 Self Diagnosis and Error M...

Page 12: ...RH Dew point Td Working range 20 80 C 4 176 F Accuracy 2 C 3 6 F for Tambient Td 20 C 36 F General Response time RH t10 90 at 20 C 68 F typ 15 s with stainless steel grid filter2 Output signal 0 1 5...

Page 13: ...lin 69130 Ecully Tel 33 4 74 72 35 82 E mail info epluse fr E E Elektronik Germany Sch ne Aussicht 8 C 61348 Bad Homburg Tel 49 6172 13881 0 E mail info epluse de E E Elektronik Italy Via Alghero 17 1...