User Manual EE1900 Humidity Module for Climate Chambers

1.2.3  Intended Use

The EE1900 humidity module is optimised for the measurement of relative humidity (RH) and dew point 

temperature (Td) in climate chambers. The use of the EE1900 other than described in this manual is 

not allowed.

The manufacturer cannot be held responsible for damages as a result of incorrect handling, installation 

and maintenance of the device.

Unauthorized modifications of the product lead to loss of all warranty claims. The device may only be 

powered with separated extra-low voltage (SELV).

Disclaimer

The manufacturer or his authorized agent can only be held liable in case of willful or gross negligence. 

In any case, the scope of liability is limited to the corresponding amount of the order issued to the 

manufacturer. The manufacturer assumes no liability for damages incurred due to failure to comply with 

the applicable regulations, operating instructions or the operating conditions. Consequential damages 

are excluded from the liability.

1.3  Environmental Aspects

Products from E+E Elektronik® are developed and manufactured observing of all relevant requirements 

with respect to environment protection. Please observe local regulations for the device disposal.

For disposal, the individual components of the device must be separated according to local recycling 

regulations. The electronics shall be disposed of correctly as electronics waste.

1.4  ESD Protection 

The sensing elements and the electronics board are ESD (electrostatic discharge) sensitive 

components of the device and must be handled as such. The failure to do so may damage the device 

by electrostatic discharges when touching exposed sensitive components.

1.5  Scope of Supply

•  EE1900 according to ordering guide

•  Inspection certificate according to DIN EN 10204-3.1

2  Product Description

2.1  General

The sensing probe and the electronics of the EE1900 undergo the factory adjustment together and may 

not be separated from each other. Do not cut, shorten or extend the probe cable. 

The analogue output of EE1900 can be set to voltage or current with a slide switche on the electronics 

board (see chapter 4.1). 

The RH output is preferred for applications with rather uniform temperature distribution, while the Td 

output is advisable in case of large T gradients in the environment. 

Summary of Contents for EE1900

Page 1: ...BA_EE1900 v1 1 Modification rights reserved User Manual EE1900 HUMIDITY MODULE FOR CLIMATE CHAMBERS...

Page 2: ...equipment has been tested and found to comply with the limits for a Class A digital device pursuant to part 15 of the FCC Rules These limits are designed to provide reasonable protection against harmf...

Page 3: ...al 5 2 1 1 Sensing Probes 6 2 1 2 Filter Caps 6 2 2 Dimensions 6 2 3 Installation 6 2 3 1 Electronics Board 6 2 3 2 Sensing Probe 6 3 Electrical Connection 7 4 Setup and Configuration 8 4 1 Selection...

Page 4: ...commissioning shall be performed by qualified personnel only 1 2 2 Mounting Start up and Operation The device has been produced under state of the art manufacturing conditions has been thoroughly test...

Page 5: ...E E Elektronik are developed and manufactured observing of all relevant requirements with respect to environment protection Please observe local regulations for the device disposal For disposal the i...

Page 6: ...47 Probe length see ordering guide Cable length see ordering guide Filter length see table below 3 5 0 14 3 5 0 14 Fig 1 Dimensions of EE1900 2 3 Installation ESD handling regulations must be strictly...

Page 7: ...tor If the probe is installed into a partition wall the probe s backend shall be isolated thermally 3 Electrical Connection Supply Fig 2 Power supply 15 35 V DC 17 29 V AC Connection supply and output...

Page 8: ...RH and error indication according NAMUR see chapter 4 3 is disabled After switching the output between voltage and current with the slide switch the scaling of the output can be set using the E E Pro...

Page 9: ...manually by pressing the Sensor Recovery button No 6 Fig 3 for min 3 seconds or automatically by applying a voltage of 5 30 V DC to the potential free input No 3 Heat IN Fig 3 for at least 3 seconds A...

Page 10: ...overy The ARC mode is indicated by the closed contact at the optocoupler output No 3 Status Fig 3 and by the red LED flashing very quickly The optocoupler output thus indicates wether the data on the...

Page 11: ...guration PC HA011017 EE1900 The adjustment shall be performed for the RH output against an appropriate RH reference See Calibration Kit User Guide at www epluse com ee1900 7 Self Diagnosis and Error M...

Page 12: ...RH Dew point Td Working range 20 80 C 4 176 F Accuracy 2 C 3 6 F for Tambient Td 20 C 36 F General Response time RH t10 90 at 20 C 68 F typ 15 s with stainless steel grid filter2 Output signal 0 1 5...

Page 13: ...lin 69130 Ecully Tel 33 4 74 72 35 82 E mail info epluse fr E E Elektronik Germany Sch ne Aussicht 8 C 61348 Bad Homburg Tel 49 6172 13881 0 E mail info epluse de E E Elektronik Italy Via Alghero 17 1...

Reviews: