20.10.
Replacing the DYSIS Acetic Acid Applicator
20.10.1.
Removing the existing Acetic Acid Applicator Kit:
1.
Push the container upwards from the bottom in order to remove it from the holder Fig 52).
2.
Hold the container and gently pull the two sides of the T-valve from its holder until the syringe
detaches from its position.
3.
Gently squeeze the pins together on the rear end of the speculum connector rod and remove it from
its mount.
Fig 52: Removing the acetic acid kit.
20.10.2.
Connecting the Acetic Acid Applicator Kit
1. Insert the speculum connector into the slot in front of the camera housing.
2. Pull the plunger of the syringe and snap into the syringe port located on the front of the rubber
housing cover (Fig 53).
3. Place the acetic acid reservoir into the holder