Dynamic DX-ARC-SWB Installation Manual Download Page 1

 

  

 

 

 

 

 

 

 

 

DX Switchbox (DX-ARC-SWB) 

Installation Manual 

 

 

 

 

by 

 

 

GBK62297 Issue 1, February 2003 

Summary of Contents for DX-ARC-SWB

Page 1: ...DX Switchbox DX ARC SWB Installation Manual by GBK62297 Issue 1 February 2003...

Page 2: ...DX ARC SWB is not user serviceable Specialized tools are necessary for the repair of any DX ARC SWB component Do not install maintain or operate this equipment without reading understanding and follo...

Page 3: ...ctions and Wiring 4 4 1 DXBUS 4 4 2 Seat Switches 5 4 3 Switched Auxiliary Power Supply Output 5 5 Testing the Installation 6 6 Specifications 6 6 1 Electrical Specifications 6 6 2 Physical Specificat...

Page 4: ...GBK62297 Issue 1 February 2003...

Page 5: ...ted to a DB15 connector Functionally it is equivalent to a DX Actuator Remote Control DX ARC5 but with the added versatility of allowing switch type and position to be tailored to suit specific indivi...

Page 6: ...ented if desired using the supplied mounting kit It is preferable that the DX ARC SWB is mounted in the following orientations Parts supplied with the DX ARC SWB Mounting bracket and screws DB15 Conne...

Page 7: ...No depending on the Master Remote used using the Wizard programmer Most DX Master Remotes will allow only one Actuator control the DX ARC SWB or the Master Remote but not both Set any other parameter...

Page 8: ...e DX ARC SWB disable the powerchair by one of the following means to prevent accidental movement 1 Place the battery circuit breaker in the open position 2 Disconnect the motor or batteries and or 3 E...

Page 9: ...ink between Pins 1 and 8 for the unit to operate Warning Use appropriate quality assemblies for the installation of the system 4 3 Switched Auxiliary Power Supply Output An unregulated 24 Volt 1 Amp m...

Page 10: ...he DX Master Remote 2 Check that the green DX ARC SWB Status light turns on and is steady i e not flashing 3 Operate each seat switch in turn and check for the correct operation of all seat functions...

Page 11: ...ter Material Aluminium alloy Finish Epoxy powder coated Rolling Thunder Protection Rating Not rated Physical Size 95mm x 61mm x 26mm Shipping Weight 100g Min Nominal Max Units Operating Temperature Ra...

Page 12: ...h Standards DX ARC SWB has been designed such that the combination of the wheelchair controller and the DX ARC SWB conforms to the requirements of ISO7176 However final conformity of the complete whee...

Page 13: ...if internal damage may have occurred for example by being dropped have it checked by qualified personnel before operating 7 3 Warranty All equipment supplied by Dynamic Controls is warranted by the c...

Page 14: ...wing the proper instructions and manuals otherwise injury or damage can result Use appropriate quality assemblies for the installation of the system 7 5 Contact Details Dynamic has a global network of...

Reviews: