Chapter 4: Programming the Rhino2
76
4.4.3.3
Forward Deceleration
Parameter
Possible Values
Default
HHP
Lite
Std Adv
Forward Deceleration
0 - 100 %
DS90, DS120:
60 % (Profile1)
60 % (Profile 2)
DS160:
65 % (Profile1)
60 % (Profile 2)
Sets how quickly the scooter slows down after the throttle has been released from forward
deflection to neutral.
0 %
- From full forward speed, the scooter stops in 10s
100 %
- From full forward speed, the scooter stops in 0.4s
Low deceleration values produce a gentle stop but increase the braking distance. High
deceleration values produce a more aggressive stop but can be uncomfortable. Adjust this
parameter to the preference of the user.
Note:
This parameter cannot be set higher than the value of the
parameter (4.4.4.6) that has been set by the scooter manufacturer.
Warning:
Setting
Forward Deceleration
too low or too high can result in a scooter that is
unsafe. Test thoroughly after programming to make sure that the scooter
complies with local regulatory requirements for maximum allowable braking
distance.
SPEED
Too low
Unresponsive
Braking distance
too long
Too high
Rough
The user can
fall out of the
scooter
Good
Smooth
Dependent on user
and environment