Maintenance
The ACCU-
CAL™ 50 was designed to operate with minimum
maintenance. Follow the schedule below to assure top unit
performance.
•
Calibrate the instrument at least annually. Calibration
service is available through Dymax Customer Service or
Dymax Product Repair.
•
Change the batteries when a low battery warning is
received. The ACCU-
CAL™ 50 uses two AA
-type
batteries. The battery compartment is on the back of the
instrument.
•
Keep the detector head’s sensing element clean and free
of contaminants. The detector head may be cleaned with
a clean tissue wetted with isopropyl alcohol.
•
To increase battery life, turn unit power off after use. The
unit will not automatically power off.
Spare Parts and Accessories
Item
Part Number
Adapter Kit
–
Flood to Spot Model (includes parts listed below)
39554
Lightguide Simulator
38408
8-mm Lightguide Adapter
39558
5-mm Lightguide Adapter
39557
3-mm Lightguide Adapter
39556
Specifications
Property
Specification
Power
Two AA size batteries
Display
LCD Graphic Display 97x 32 Pixel
Display area: 0.56 in x 1.41 in (14.3 mm x 35.8 mm)
Detector Interface
9-Pin MDSM9 socket, 4 inputs
Measurement
Ranges
Four modes of operation
Auto range within each operating mode
Front Panel Control
3 buttons
Temperature
Operating: 5 to 40°C
Storage: -10 to 50° C
Size
5.71 in x 2.48 in x 1.18 in (145 mm x 63 mm x 30 mm)
Weight
0.33 lbs (150 g)
Figure 7.
Battery Compartment (Closed & Open)