Dwyer Instruments PMT2 Series Installation And Operating Instructions Manual Download Page 1

Series PMT2 Particulate Transmitter

Specifications - Installation and Operating Instructions

Bulletin PC-PMT2-M

2X 1/2˝ NPT OR M20 ELECTRICAL ENTRY

5/8

[15.88]

“A”

1-1/2

[38.10]

4-15/16

[125.41]

3/4

[19.05]

3-29/32

[99.22]

3/4˝ NPT or 3/4˝ BSPT

ALLOW 1-1/4˝ [32]

FOR COVER REMOVAL

BOTH SIDES

4-3/8

[111.13]

2X 1/2˝ NPT OR M20 ELECTRICAL ENTRY

1˝ NPT

“A”

“B”

5-13/16

[147.64]

3-29/32

[99.22]

THREADED MOUNT

Probe

Length

A

in [mm]

10˝

15˝

20˝

30˝

36˝

3 [76.20]

5 [127.00]

10 [254.00]

15 [381.00]

20 [508.00]

30 [762.00]

36 [914.40]

1.5˝ TRI-CLAMP MOUNT

Probe

Length

A

in [mm]

B

in [mm]

10˝

15˝

20˝

30˝

36˝

3-1/2 [88.90]

5-1/2 [139.70]

10-1/2 [266.70]

15-1/2 [393.70]

20-1/2 [520.70]

30-1/2 [744.70]

36-1/2 [927.10]

1-25/32 [45.24]

3-25/32 [96.04]

8-25/32 [223.04]

13-25/32 [350.04]

18-25/32 [477.04]

28-25/32 [731.04]

34-25/32 [883.44]

1.5˝ Tri-Clamp Option Mount

3/4˝ Male NPT OR 3/4˝ Male BSPT Mount

The 

SERIES PMT2 

Particulate Transmitter is designed to measure particulate emission 

levels from dust collector discharge. Using DC coupled electrostatic induction sensing 

technology, the transmitter monitors a pA current that is generated as particulate 

passes near the probe; a 4 to 20 mA signal will vary based on the particulate level. The 

PMT2 offers 6 sensitivity ranges allowing the user to choose the range that will best fit 

the application. The range and test selector switch can also be set to output a 4 mA or 

20 mA signal to assist with set up or trouble shooting. Averaging time setting can be 

used to dampen the signal if desired. 

FEATURES/BENEFITS

•  Simple 2-wire installation for PLC and control panels

•  Non-stick PTFE coated probe to prevent false readings from moist and conductive 

  dusts, condensate, and dust buildup

•  Remote zero calibration helps to decrease maintenance time

SPECIFICATIONS

Service:

 Air and compatible gases, any type of particulate conductive or non-

conductive.

Wetted Materials: 

316L SS, silicone, and PTFE.

Enclosure: 

Powder coated aluminum.

Accuracy: 

±5% of reading.

Particulate Size: 

0.3 microns and higher.

Detection Range: 

5 to 5000 pA (6 selectable range options).

Temperature Limits: 

Ambient: -40 to 145°F (-40 to 63°C); Process: -40 to 248°F 

(-40 to 120°C).

Pressure Limit: 

30 psi (2 bar).

Output Signal: 

4 to 20 mA.

Power Requirements: 

12 to 28 VDC (    ).

Electrical Connection: 

Two 1/2˝ female NPT electrical entries or two M20 electrical 

entries (A2 suffix only).

Terminal Block:

 Removable (16 to 20 AWG wire).

Process Connection: 

See model chart. BSPT process connections are not UL 

listed.

Probe Lengths: 

See model chart.

Enclosure Rating: 

UL Type 4 (IP66) ATEX/IECEx IP65.

Mounting Orientation:

 Any.

Averaging Time: 

1 to 360 s (10 selectable options).

Weight: 

Varies with length of probe and type of mount. 

Agency Approvals: 

CE, cULus; ATEX Compliant:     0518     II 1 G Ex ia IIB T4 GA 

(-40°C ≤ Tamb ≤ 63°C) (-40°C ≤ T 

Process 

≤ 120°C) / II 1 D Ex ia IIIC T120°C Da

(-40°C ≤ Tamb ≤ 63°C) (-40°C ≤ T 

Process 

≤ 120°C). Type Certificate No.: DEMKO 

16ATEX1768 X. ATEX Standards: EN 60079-0:2012/A11:2013; EN 60079-11:2012. 

IECEx Certified: Ex ia IIB T4 Ga (-40°C ≤ Tamb ≤ 63°C) (-40°C ≤ T 

Process 

≤ 120°C) / 

Ex ia IIIC T120°C Da (-40°C ≤ Tamb ≤ 63°C) (-40°C ≤ T 

Process 

≤ 120°C).

Certificate of Conformity: IECEx UL 16.013X. IECEx Standards: IEC 60079-0: 2011; 

IEC 60079-11: 2011. UL Listed Intrinsically Safe for Class I, Groups C and D; Class 

II, Groups E, F and G; Class III; Class I Zone 0 AEx ia IIB T4 Ga; Class I Zone 0 Ex 

ia IIB T4 Ga.

MODEL CHART

Example

PMT2 -05 -A -U2

PMT2-05-A-U2

Series

PMT2

Particulate transmitter

Probe

Length

03

05

10

15

20

30

36

3˝ probe length

5˝ probe length

10˝ probe length

15˝ probe length

20˝ probe length

30˝ probe length

36˝ probe length

Process

Connection

A

B

C

3/4˝ male NPT

1.5˝ tri-clamp kit with 1˝ male NPT

3/4˝ male BSPT

Enclosure Rating

A2

U2

ATEX and IECEx (IS)

UL (IS)*

Options

ST

M2

Stainless steel tag

Female M20 electrical entries

(female 1/2˝ NPT standard)

*Options that do not have ATEX or IECEx.

Attention: Units without the A2 suffix are not Directive 2014/34/EC (ATEX) 

compliant. These Units are not intended for use in potentially hazardous 

atmospheres in the EU. These unites may be CE marked for other Directives 

of the EU. 

DWYER INSTRUMENTS, INC.

P.O. BOX 373 • MICHIGAN CITY, INDIANA 46360, U.S.A.

Phone: 219/879-8000
Fax: 219/872-9057

www.dwyer-inst.com
e-mail: [email protected]

®

Summary of Contents for PMT2 Series

Page 1: ...Range 5 to 5000 pA 6 selectable range options Temperature Limits Ambient 40 to 145 F 40 to 63 C Process 40 to 248 F 40 to 120 C Pressure Limit 30 psi 2 bar Output Signal 4 to 20 mA Power Requirements...

Page 2: ...ity ranges allow the PMT2 to monitor particulates with low charge properties or high charge properties As a reference Table 1 lists particulate charge properties and the suggested range INSTALLATION U...

Page 3: ...o after a filter failure or filter changes Re zeroing should only be done when there is no air flow in the duct Averaging Time Selection Switch The PMT2 will average the output for the selected amount...

Page 4: ...ticulate transmitter It is suggested to set two alarm set points One alarm set point to monitor the emission spikes and the second alarm to detect an increase in the baseline The alarm monitoring the...

Page 5: ...ile the transmitter is zeroing the PMT2 will output about 3 5 mA The zero function will take approximately 3 minutes When zeroing is complete the output will return to a normal output signal and the t...

Page 6: ......

Page 7: ...DWYER INSTRUMENTS INC P O BOX 373 MICHIGAN CITY INDIANA 46360 U S A Phone 219 879 8000 Fax 219 872 9057 www dwyer inst com e mail info dwyermail com...

Page 8: ...chtetes Aluminium Genauigkeit 5 des Messwerts Partikelgr e 0 3 Mikrometer und mehr Messbereich 5 bis 5000 pA 6 ausw hlbare Messbereichsoptionen Temperaturbereich Umgebungstemp 40 to 145 F 40 bis 63 C...

Page 9: ...mpfindlicheitsstufen lassen sich mit dem PMT2 sowohl Partikel mit niedriger als auch mit hoher Ladung berwachen Zur Orientierung sind in Tabelle 1 die Ladungseigenschaften und die empfohlenen Stufen a...

Page 10: ...ullabgleich nur dann durchf hren wenn keine Luft durch die Leitung str mt Auswahlschalter Mittelungszeit Der PMT2 mittelt das Ausgangssignal ber den angegebenen Zeitraum So werden Spitzen ausgeglichen...

Page 11: ...rde Alarme k nnen in einer SPS oder Steuerungsanlage auf der Grundlage des Signals von 4 bis 20 mA aus dem Partikeltransmitter programmiert werden Es empfiehlt sich zwei Alarmsollwerte einzustellen Da...

Page 12: ...er Nullabgleich dauert circa 3 Minuten Nach Abschluss des Nullabgleichs normalisiert sich das Ausgangssignal und der Transmitter ist einsatzbereit Abbildung 4 Nullabgleich aus der Ferne und Nullschalt...

Page 13: ......

Page 14: ...DWYER INSTRUMENTS INC P O BOX 373 MICHIGAN CITY INDIANA 46360 USA Tel 1 219 879 8000 Fax 1 219 872 9057 www dwyer inst com E Mail info dwyermail com...

Page 15: ...tions de plage possibles Limites de temp rature ambiante 40 145 F 40 63 C en fonction 40 248 F 40 120 C Limite de pression 2 psi 30 bar SIgnal en sortie 4 20 mA Puissance lectrique n cessaire 12 28 VC...

Page 16: ...Les six plages de sensibilit permettent au PMT2 de surveiller aussi bien des particules charge basse que des particules charge haute titre de r f rence le tableau 1 liste les charges des particules et...

Page 17: ...la conduite Commutateur de s lection de dur e moyenne Le PMT2 fait la moyenne des niveaux en sortie sur la dur e s lectionn e Cela aura pour effet de lisser les cr tes du signal apparues pendant les...

Page 18: ...sonde metteur particules Il est recommand de param trer deux points d alerte Un point d alerte pour surveiller les pics d mission et le second pour d tecter une augmentation par rapport au niveau de r...

Page 19: ...tialise le PMT2 enverra environ 3 5 mA La fonction de r initialisation dure environ 3 minutes Lorsque la r initialisation est termin e le signal redevient normal et l metteur est pr t fonctionner Figu...

Page 20: ......

Page 21: ...DWYER INSTRUMENTS INC P O BOX 373 MICHIGAN CITY INDIANA 46360 tats Unis T l phone 1 219 879 8000 Fax 1 219 872 9057 www dwyer inst com courriel info dwyermail com...

Page 22: ...to 5000 pA 6 opciones de rango elegibles L mites de temperatura Ambiente 40 a 145 F 40 a 63 C proceso 40 a 248 F 40 a 120 C L mite de presi n 30 bar 2 psi Se al de salida 4 a 20 mA Requisitos de alim...

Page 23: ...e Los seis rangos de sensibilidad permiten que el PMT2 monitoree part culas con propiedades de baja carga o de alta carga Como referencia la tabla 1 presenta las propiedades de carga de las part culas...

Page 24: ...n el conducto Interruptor de selecci n de tiempo de promedio El PMT2 promediar la salida durante el tiempo seleccionado Esto amortiguar los picos en la salida causados durante los ciclos de limpieza n...

Page 25: ...ulas Se recomienda ajustar dos puntos de alarma Un punto de ajuste de alarma para monitorear los picos de las emisiones y la segunda alarma para detectar un incremento en la base de referencia La alar...

Page 26: ...n de calibraci n a cero tardar unos 3 minutos Al finalizar la calibraci n a cero la salida regresar a una se al de salida normal y el transmisor estar listo para operar Figurae 4 cero remoto e interru...

Page 27: ......

Page 28: ...truments Inc Printed in U S A 5 17 FR 444122 10 DWYER INSTRUMENTS INC P O BOX 373 MICHIGAN CITY INDIANA 46360 ESTADOS UNIDOS Tel fono 1 219 879 8000 Fax 1 219 872 9057 www dwyer inst com Correo electr...

Reviews: