background image

1  WATER PROBE SETUP 

1.1  Connections 

1.1.1  Air Loop 

Two pieces of tubing connect the RAD7 and drying 
unit to the Water Probe air/water exchanger, as 
shown in Fig. 2. 

Th

ese two pieces of tubing can be 

several tens of meters long. 

Th

e standard tubing 

supplied with the RAD7/Water Probe is su

cient for 

a connection up to ten feet between the exchanger 
and the RAD7. 

Connect the OUTLET of the RAD7 to either of the 
Water Probe air ports. For this, the 5

ft

 long, 3/16" ID 

tubing, with a 1/8" ID section at one end, may be 
used. 

Th

e 1/8" end 

ts on the RAD7 outlet, and the 

3/16" end 

ts the Water Probe. Connect the other 

3/16” hose connector between the other Water Probe 
air port and the Laboratory Drying Unit. 

Connect the other end of the Laboratory Drying Unit 
(there should be at least one inch of blue desiccant 
le

ft

 at this end) to the air inlet 

lter (with 1/8" ID 

tubing at the 

lter end), which is then placed on the 

RAD7 INLET. 

Th

e Luer taper ensures an airtight 

connection. 

 

Fig. 2  RAD7 Water Probe Standard Setup 

1.1.2  DRYSTIK 

Please note that the above instructions are for use of 
the Water Probe without a DRYSTIK humidity 
exchanger. A DRYSTIK, if available, should be placed 
between the exchanger and the drying unit, and the 
outer sheath should be purged with dry air from the 
RAD7 outlet. See Fig. 3 in Section 3. More details are 
provided in the DRYSTIK user’s manual. 

1.1.3  RAD7 and Exchanger Location 

Place the RAD7 on a clean, dry surface. If it has to be 
located in a harsh environment, then it should be 
protected from the elements (especially water). A 
simple way to do this is to place the RAD7 inside a 
large transparent plastic bag, such as the one in which 
it was originally shipped. 

Th

e bag opening should be 

gathered around the inlet and outlet tubes, so that the 
instrument is inside a closed space, completely 

Introduction

5

Summary of Contents for RAD7

Page 1: ...WATER PROBE Passive Slow Response Radon in Water Accessory for the RAD7 User Manual...

Page 2: ...Protocol 6 1 4 3 User Protocol 6 2 MEASUREMENT PROCEDURE 7 2 1 Start Up 7 2 1 1 Temperature Probe 7 2 1 2 Start Measurement 7 2 2 Speed of Response 7 2 2 1 Measurement in Progress 7 2 2 2 In fl uenci...

Page 3: ...ff ect on Response Time 11 4 4 Custom DRYSTIK Settings 11 Fig 4 Water Probe con fi guration with Active DRYSTIK 12 5 CARE MAINTENANCE AND TROUBLESHOOTING 13 5 1 Water Catastrophe 13 5 2 RAD7 Care 13...

Page 4: ...mperature is measured with a temperature probe inserted into the water As an example at a water temperature of 10 degrees Celsius the partition coe ffi cient is about 1 3 Th at means there is three ti...

Page 5: ...t this end to the air inlet fi lter with 1 8 ID tubing at the fi lter end which is then placed on the RAD7 INLET Th e Luer taper ensures an airtight connection Fig 2 RAD7 Water Probe Standard Setup 1...

Page 6: ...tteries in a fully charged condition Set the RAD7 s Mode to Auto which starts the instrument in Sni ff mode counting 218 polonium decays only and switched automatically to Normal Mode counting 218 pol...

Page 7: ...a standard deviation of 10 At very low concentrations it may take hours and averaging over many cycles to reach a su ffi ciently precise value 2 2 2 In fl uencing Factors Th ere are two processes req...

Page 8: ...the process and which must be replenished Th e most obvious is the desiccant A new or regenerated Laboratory Drying Unit will normally last for about ten days of continuous use in a temperate climate...

Page 9: ...ce as instructed Th e water temperature information is used by CAPTURE to calculate the radon in water concentrations For detailed instructions please see the Water Probe information in the CAPTURE us...

Page 10: ...formula is applied automatically in CAPTURE when the Water Type is set to Fresh Water in the Run Parameters Window as shown in Fig 3 below 3 2 2 Schubert Et Al Formula If the Water Type is set to Sal...

Page 11: ...It should be inserted between the DRYSTIK air outlet and the inlet fi lter on the RAD7 as shown in Fig 4 on the following page Th e desiccant will last for a very long time and the drying tube will ad...

Page 12: ...Fig 4 Water Probe con fi guration with Active DRYSTIK Section 4 DRYSTIK 12...

Page 13: ...id onto its hinges Th e wrap will make the RAD7 almost watertight If it tears it can be easily replaced at any time Th e instrument should in any case be returned every year for recalibration It is us...

Page 14: ...dioxide measurements in coastal waters Environmental Science and Technology 46 14 7685 7691 Schubert M Paschke A Lieberman E Burnett W C 2012 Air Water Partitioning of 222Rn and its Dependence on Wat...

Page 15: ...978 667 9556 Fax 978 667 9557 Web www durridge com Email service durridge com Copyright 2021 DURRIDGE Company Inc All rights reserved DURRIDGE the DURRIDGE logo and the Works with the DURRIDGE logo ar...

Reviews: