background image

5  CARE, MAINTENANCE, AND TROUBLESHOOTING 

5.1  Water Catastrophe 

If water ever enters the RAD7, or if the RAD7 ever 
goes swimming in the water, it will probably cease to 
operate and immediate steps should be taken to 
minimize the impact on the instrument.  

Keep the RAD7 upright. 

Th

is will prevent water from 

touching the detector, which is close to the face plate 
at the top of the dome. Put a piece of tubing on the 
RAD7 outlet with the other end in a sink. Use the 
RAD7 pump if it still works or, otherwise, an external 
pump into the inlet, to blow air through the 
instrument. When water ceases to be blown out of 
the outlet, put desiccant upstream of the RAD7 to dry 
out the air path. When the air path is fully dry (a

ft

er 

dry air has been blown through it for approximately 
one hour), remove the face plate from the case, empty 
the water out of the case and blow dry the case and 
the RAD7 electronics. 

Once there is no visible water in or on the 
instrument, it can be put in an oven at 50°C for a few 
hours to dry out completely. Additionally, desiccated 
air can be passed through the air path until the air 
leaving the RAD7 drops below 10% RH. A

ft

er this 

treatment further corrosion will be prevented, and 
the RAD7 will boot once more and you can use the 
internal RH sensor to measure how dry the air path 
is. At this point the instrument should be returned to 
DURRIDGE for service. 

5.2  RAD7 Care 

Water, particularly salt water, is hostile to electronic 
instruments. Please keep the RAD7 in a relatively 
clean and dry environment. One way is to enclose the 
instrument in a large, transparent plastic bag; see 
Section 1.2.3. Should it ever be seriously splashed 
with salt water, please follow the instructions in 
Section 7.1, above. 

As a preventive measure, plastic cling wrap can be 
placed over the RAD7 face plate and down the sides 
of the RAD7. Push it down around the hose 
connections, push the power and RS232 plugs into 
their sockets, and push the lid onto its hinges. 

Th

wrap will make the RAD7 almost watertight. If it 
tears it can be easily replaced at any time. 

Th

instrument should, in any case, be returned every 
year for recalibration. 

It is useful to look at a cumulative spectrum 
periodically. 

Th

is may be obtained by having the 

printer on and allowing the RAD7 to complete a run. 

Th

e “Recycle” number may be set to the current cycle 

number (

Setup, Recycle, NN

 

[ENTER]

). When 

the RAD7 reaches the end of the current cycle it will 
then print out the end of run summary including the 
cumulative spectrum. Look to see that the peaks are 
clean and in their normal positions. 

5.3  Exchanger Care 

Th

ere is a potential for biofouling to occur in long-

term monitoring applications. 

Th

e exchanger should 

be kept as clean as possible in the circumstances. 

Th

Water Probe should therefore be inspected 
periodically to ensure that there is good contact 
between the radon exchange tubing and the 
surrounding water, and no clogging with weeds, 
algae, etc. 

5.4  Desiccant Regeneration 

Please see the RAD7 manual for information on the 
care and regeneration of the desiccant. Regenerated 
desiccant, a

ft

er a few regenerations, loses most of its 

indicating ability (due, we believe, to migration of the 
cobalt chloride to the interior of the calcium sulphate 
crystals). One way to ‘indicate’ the status is, every 
time you re

ll the laboratory drying unit with 

regenerated desiccant, you 

rst add half an inch or so 

of new, blue desiccant, out of the jar. 

Th

is way, you 

can always tell if the unit is still working, as the new 
desiccant will only turn pink when the rest of the 
desiccant, upstream, has become hydrated. 

5.5 Water Probe Troubleshooting 

5.5.1  Air Path Integrity 

When drawing a sample from a remote location, air 
path integrity is essential to prevent dilution of the 
sample with ambient air. Always make sure that there 
are no loose connections or leaky 

ttings (such as the 

screw cap of the Laboratory Drying Unit) in the air 
loop, particularly upstream of the RAD7. In the event 
of unexpectedly low radon values, check the air path 
for integrity. 

Section 5 

 

Care, Maintenance, and Troubleshooting

13

Summary of Contents for RAD7

Page 1: ...WATER PROBE Passive Slow Response Radon in Water Accessory for the RAD7 User Manual...

Page 2: ...Protocol 6 1 4 3 User Protocol 6 2 MEASUREMENT PROCEDURE 7 2 1 Start Up 7 2 1 1 Temperature Probe 7 2 1 2 Start Measurement 7 2 2 Speed of Response 7 2 2 1 Measurement in Progress 7 2 2 2 In fl uenci...

Page 3: ...ff ect on Response Time 11 4 4 Custom DRYSTIK Settings 11 Fig 4 Water Probe con fi guration with Active DRYSTIK 12 5 CARE MAINTENANCE AND TROUBLESHOOTING 13 5 1 Water Catastrophe 13 5 2 RAD7 Care 13...

Page 4: ...mperature is measured with a temperature probe inserted into the water As an example at a water temperature of 10 degrees Celsius the partition coe ffi cient is about 1 3 Th at means there is three ti...

Page 5: ...t this end to the air inlet fi lter with 1 8 ID tubing at the fi lter end which is then placed on the RAD7 INLET Th e Luer taper ensures an airtight connection Fig 2 RAD7 Water Probe Standard Setup 1...

Page 6: ...tteries in a fully charged condition Set the RAD7 s Mode to Auto which starts the instrument in Sni ff mode counting 218 polonium decays only and switched automatically to Normal Mode counting 218 pol...

Page 7: ...a standard deviation of 10 At very low concentrations it may take hours and averaging over many cycles to reach a su ffi ciently precise value 2 2 2 In fl uencing Factors Th ere are two processes req...

Page 8: ...the process and which must be replenished Th e most obvious is the desiccant A new or regenerated Laboratory Drying Unit will normally last for about ten days of continuous use in a temperate climate...

Page 9: ...ce as instructed Th e water temperature information is used by CAPTURE to calculate the radon in water concentrations For detailed instructions please see the Water Probe information in the CAPTURE us...

Page 10: ...formula is applied automatically in CAPTURE when the Water Type is set to Fresh Water in the Run Parameters Window as shown in Fig 3 below 3 2 2 Schubert Et Al Formula If the Water Type is set to Sal...

Page 11: ...It should be inserted between the DRYSTIK air outlet and the inlet fi lter on the RAD7 as shown in Fig 4 on the following page Th e desiccant will last for a very long time and the drying tube will ad...

Page 12: ...Fig 4 Water Probe con fi guration with Active DRYSTIK Section 4 DRYSTIK 12...

Page 13: ...id onto its hinges Th e wrap will make the RAD7 almost watertight If it tears it can be easily replaced at any time Th e instrument should in any case be returned every year for recalibration It is us...

Page 14: ...dioxide measurements in coastal waters Environmental Science and Technology 46 14 7685 7691 Schubert M Paschke A Lieberman E Burnett W C 2012 Air Water Partitioning of 222Rn and its Dependence on Wat...

Page 15: ...978 667 9556 Fax 978 667 9557 Web www durridge com Email service durridge com Copyright 2021 DURRIDGE Company Inc All rights reserved DURRIDGE the DURRIDGE logo and the Works with the DURRIDGE logo ar...

Reviews: