24
2142100002L02 1908V007
Assembly
i
Insert the plug into the power outlet.
i
Refit the cover.
The cover on the rear must be correctly
fitted when the device is operated within
the patient environment.
7.7 Connecting the device to the
network
Purpose of the network connection
The network connection is used to exchange in-
formation or control signals between the unit
and a software installed on a computer, in order
to, e. g.:
– Display parameters
– Select operating modes
– Indicate messages and error situations
– Change unit settings
– Activate test functions
– Transmit data for archiving
– Provide documents concerning the units
The unit can be connected to the network with
a network cable or via WLAN.
For information on connection via WLAN
see "8.1 Installing and configuring the
unit".
Combining devices safely
– The overall safety of the unit and its main
performance features are independent of the
network. The device is designed for operation
independent of a network. However, some of
the functions are not available in this case.
– Incorrect manual configuration can lead to
significant network problems. The expert
knowledge of a network administrator is
required for configuration.
– The data connection utilises part of the band-
width of the network. Interactions with other
medical devices cannot be completely ruled
out. Apply the IEC 80001-1 standard for risk
assessment.
– The device is not suitable for direct connec-
tion to the public internet.
Take care when connecting units together or to
parts of other systems as there is always an ele-
ment of risk (e.g. due to leakage currents).
i
Only connect units when there can be no
question of danger to operator or to patient.
i
Only connect units when it is safe to do so
and when there is no risk of damage or harm
to the surroundings.
– Mains voltage must match the information
shown on the type plate of the power supply
unit
i
Attach the matching country-specific adapter
to the power supply unit.
i
Remove the cover from the rear of the device.
i
Plug in the connecting plug of the power sup-
ply unit into the socket connection of the de-
vice.
i
Secure the cable with a cable clip.
2
1
EN