Durr Dental VISTACAM2 Installation And Operating Instructions Manual Download Page 1

25

INSTALLATION AND OPERATING INSTRUCTIONS

DÜRR VISTACAM2

2000/05

Summary of Contents for VISTACAM2

Page 1: ...25 INSTALLATION AND OPERATING INSTRUCTIONS DÜRR VISTACAM2 2000 05 ...

Page 2: ...26 ...

Page 3: ...ng 32 7 1 Room requirements 32 7 2 Choice of mounting 32 7 3 Installation 32 7 4 Connecting Image Viewer 33 7 5 Handset holder 33 8 Electrical Connection 34 8 1 Video Setup 34 8 2 PC Setup 34 8 3 Foot switch 35 8 4 Power supply 35 9 Putting into Operation 36 9 1 Functional Check 36 9 2 IEC 601 Check 36 Use 10 Operation 37 10 1 IFC Optic Sheath 37 10 2 Operation Model Professional 39 10 3 Operation...

Page 4: ...authorized by Dürr Dental and if the equipment is used in conformity with the Installation and Operating Instructions These Installation and Operating Instructions conform to the relevant version of the equipment and the underlying safety standards valid at the time of going to press All switches processes trade marks software programs and appliances named in this document are registered names Any...

Page 5: ...the power supply Check the appliance and the power supply cables for possible damage before switching on Damaged cables plugs and sockets must be replaced before use Never touch open supply outlets and patients simultaneously 1 7 Warnings and Symbols In the Installation and Operating Instructions use is made of the following terms or symbols to denote information of special importance Information ...

Page 6: ... Cleaning set for Optic 2101 025 50 Disinfection wipes Dürr FD 350 Mini box XX02 94 010 2 PRODUCT INFORMATION 2 1 Correct Usage The Dürr VistaCam2 is solely designed for the production of intra and extraoral images in dental medical practices and clinics Any other use or use beyond what is specified is deemed to be not for the intended purpose The manufacturer accepts no liability for damage resul...

Page 7: ...CE Label Produced according to ICE 601 1 and MPG Basicsystem Video color system PAL NTSC Automatic brightness regulation White balance Factory set at ca 3400K Ports Video BNC S Video 4 pin Mini DIN Handset Diameter ca 2 2 cm Weight ca 65g Rapid release IFC Optik CameraelectronicProfessional Imagesensor 1 3 Color Interline Transfer CCD Pixels 320 000 Pixels effective 500 H x 582 V NTSC 510 H x 492 ...

Page 8: ...32 5 FUNCTIONAL DESCRIPTION ...

Page 9: ...eo monitor video printer video monitor PC with video board monitor conforming to section 1 5 Using peripheral devices The light is transmitted from a halogen cold light source in the base unit via a light carrier in the cable to the IFC Optic The white balance and the color control of the appliance regardless of the light setting are set at 3400K In order to achieve the optimal colour recording th...

Page 10: ...2 on either a wall or a roof column Special mounting sets can be ordered from Dürr Dental for this purpose 7 3 Installation Unpack all parts Place the appliance in position The VistaCam2 must be placed on a flat working surface The VistaCam2 must not be placed near basins or spittoons No spray water should reach the appliance When mounting on a wall or column the electrical connections must be poi...

Page 11: ...please consult your specialist supplier 7 5 Handset holder The handset holder 1 can be mounted directly on the VistaCam2 on the right or left side or mounted in any position near the VistaCam2 camera cable is 1 65 m in length Use mounting set order no 2103 105 00 The mounting of the holder is described in the installation instructions supplied with the mounting kit Please observe instructions ...

Page 12: ...ignal Please observe the installation instructions for the video printer and or video monitor 8 2 PC Setup Connect the Video BNC cable 10 to the rear of the appliance BNC connection 7 and to the video digitalization board connection incoming signal or Connect the S Video cable 4 to the rear of the appliance 4 pin MINI DIN connection 6 and to the video digitalization board connection incoming signa...

Page 13: ...0 725 99 01 8 4 Power supply The Dürr VistaCam2 is designed for use with voltage of 230V 50 60Hz Check that the required voltage is available before connecting the appliance to the power supply Failure to do so may result in damage to the equipment Only use the power cable supplied by Dürr Plug the power cable 11 in on the rear of the appliance power socket 9 and to the mains supply ...

Page 14: ...er delivery The appliance must first reach room temperature otherwise humidity and condensation can build up which could destroy the electronics when the unit is switched on Avoid any action that might endanger appliance or personnel The appliance may only be used in a safe and fully functional condition Before use check that mounting and cables are correct Switch on appliance power switch Green L...

Page 15: ...be worn before removing the disposable sheaths from their packaging Hold the handset so that the optic is face down Hold the packed sheath so that the plastic side of the packaging together with the labeling is face up Slightly lift the white edge of the sheath and carefully slide the IFC optic in Once the tip of the optic is as far in as possible stretch the sheath 2 3 mm further so that the shea...

Page 16: ...FC Optic Carefully pull off the paper packaging in the direction of the handset The optical window in the sheath should lie exactly over the optical window of the IFC Optic After use peel the sheath off carefully and dispose of as appropriate Careless removal of the sheath the IFC Optic may be pulled from the handset and could fall to the floor and be damaged Extraoral images can be taken without ...

Page 17: ... the IFC Optic As soon as the handset is removed from its holder the integrated lighting is automatically switched on The appliance is ready for use If the integrated lighting is not required this can be switched off on the unit Switch off the lighting after each use i e replace the handset into the holder This will both protect and improve the life of the high quality lamp In cases of high contra...

Page 18: ...After imaging the handset must be disinfected Place handset in holder The handset should never be placed on the unit danger of rolling off or between treatment instrumentation the IFC Optic can be damaged by sharp metallic objects For extraoral images it is advisable to switch off the integrated lighting at the switch and or to take images without the sheath This prevents milky images Only clean t...

Page 19: ...ges can be individually altered or reloaded using the left panel 15 The image position from 1 to 4 is shown by the white lines displayed top left on the screen Pressing the middle panel 16 or turning the appliance off deletes all images 11 CLEANING Remove the sheath after every patient 11 1 Cleaning the Optic Remove the optic from the handset We recommend to use the Cleaning Set for VistaCam Optic...

Page 20: ...orrectly Turn off the machine at the power switch Unplug from the mains supply Place handset in holder For models with foot switch Freeze Module unplug the foot switch from the appliance or PC Push the trolley to the appropriate treatment room connect foot switch if relevant plug unit into the mains supply During transport take care that the cables do not get caught e g cables are not pulled out o...

Page 21: ...ool or use a cloth or gloves when changing the lamp Turn off the appliance at the power switch and unplug Turn the appliance round the front should be forwards On the bottom of the appliance is a flap closed with a spring catch Slide the spring catch in the direction of the flap and open Extract the lamp with holder remove the holder from the lamp Only use original lamps as they guarantee maximum ...

Page 22: ...ook to the video signal digital board Framegrabber of the Computer section Frame Field TROUBLESHOOTING 14 TIPS FOR OPERATOR 1 Image opaque milky Sheath incorrectly positioned IFC Optic dirty Correctly align position disposable sheath see section 10 Operation Clean the IFC Optic see section 11 Cleaning Press switch on appliance Change the lamp see section 13 1 Lamp Change See section 13 1 Lamp Chan...

Page 23: ...AL 15 DISPOSAL OF APPLIANCE All electronic components including boards should be disposed of as electronic waste All other parts can be disposed of according to the relevant regulations concerning waste disposal ...