Durr Dental VC 45 Installation And Operating Instructions Manual Download Page 1

EN

9000-606-70/30



0297

VC 45 Surgical Suction Unit

Installation and Operating Instructions

2016/11SE

Summary of Contents for VC 45

Page 1: ...EN 9000 606 70 30 0297 VC 45 Surgical Suction Unit Installation and Operating Instructions 2016 11SE ...

Page 2: ......

Page 3: ... functioning 24 16 6 Storage 24 16 7 Disinfection and cleaning of the appliance surfaces 24 17 Maintenance 25 18 Fuse changing 25 Troubleshooting 19 Tips for Operators and Service Technicians 26 Important Information 1 General 4 1 1 Evaluation of conformity 4 1 2 General notes 4 1 3 Appliance disposal 4 1 4 Notes on Medical Products 4 1 5 Notes on EMC for medical products 5 1 6 Correct Usage 5 1 7...

Page 4: ...ant version of the equip ment and the underlying safety standards val id at the time of going to press All circuits processes names software and appliances quoted are protected under industrial property rights The translation of these Installation and Oper ating Instructions has been carried out in good faith No liability can be accepted for mistakes in the translation The enclosed Ger man version...

Page 5: ...le in jury or danger In spite of this we feel it is our duty to mention the following safety measures in order to prevent any possible danger When operating the appliance observe all lo cal rules and regulations Converting or modifying the appliance in any way is strictly prohibited In such cases any and all guarantees immediately become invalid The operation of modified appliances can be punishab...

Page 6: ... instructions or prohi bitions regarding personal safety or ex tensive material damage Warning dangerous electrical voltage Steam sterilise at 134 C Comply with the operating instructions 3 1 Model identification plate The model identification plate can be found on the underside of the appliance near the power supply connection Order no Type no Serial no Batch code Observe accompanying documentati...

Page 7: ...Special accessories The following are optional and can be used with the unit VC 45 Cart 0670 150 00 Collector for disposable bags 0670 980 05E Surgical cannula ø 2 5 mm 0700 007 50 Hose adapter 6 mm 5 pieces 0670 981 03 4 4 Disposable materials The following materials are required when oper ating the unit and must be ordered separately Bacteria filter 10 pieces 0670 982 00 Suction hose silicone ca...

Page 8: ...s and other particles Within the collector lid there is a bacteria filter which is de signed so that fluids are prevented from flowing to the vacuum pump Below this bacteria filter there is a spray shield which stops any fluids being splashed towards the bacteria filter Once work has ceased the unit should be switched off at the main power supply switch Before removing the collector vessel in orde...

Page 9: ... due to build up of condensation 7 2 Set up alternatives Set up the VC 45 on a stable flat and raised surface next to the treatment unit Operation of the VC 45 directly on the floor is not permitted Set up the VC on the special cart available as special accessory 2 8 Electrical Connection Before first use during commissioning check that the supply voltage is the same as that given on the model ide...

Page 10: ... surface Press the cover carefully down using both hands until the cover engages The bacteria filter excess suction stop is only to be used once and must be changed at least once a day If the unit is to be used continuously for longer periods several hours it is im portant to change the bacteria filter after each patient Place the bacteria filter 15 onto the mould ings of the collector vessel hand...

Page 11: ...til it locks into place Slide the suction hose 18 onto its adapter 19 Plug the hose adapter into the opening of the collector lid slightly turning it to increase the locking effect 6 7 8 Clip the cannula retainer 20 onto the front of suction handpiece 21 Fit the O ring 22 onto the rear of suction handpiece Press the suction handpiece and hose adapt er piece 23 together Slide the suction hose 18 on...

Page 12: ...uction handpiece This cannula retainer is designed for cannulae up to 6 mm in diameter 10 Operating the appliance Before treating a patient we strongly recom mend that two capfuls approx 40 ml of Orotol Plus or any other agent specifically approved by Dürr Dental be poured into the collector vessel 10 1 Switch the appliance On Off Appliance switch has LEDs 25 To switch on the appliance press switc...

Page 13: ...nula have been sterilised During aspiration observe the level of fluid in the collector vessel The bacteria filter excess vacuum stop prevent fluids from getting into the pump Nevertheless the collector should be changed when the fluid level reaches approx 1 litre 10 4 Checking the bacteria filter excess vacuum stop The bacteria filter excess suction stop is only to be used once and must be change...

Page 14: ...rs 13 40 41 49 Place the securing clip 44 onto the adapter The collector should now hang in the adapter guide Before every application carry out a vis ual check of the complete appliance in cluding the collector If the collector shows any signs of fatigue small cracks because of frequent disinfection or sterilisation it should be replaced as soon as possible 14 44 on the vacuum setting with an ope...

Page 15: ...nd simultaneously apply slight pressure to the middle of the cover of the disposable bag Briefly wait until the bag has completely unfolded Slide the suction hose 51 onto the suction handpiece and elbow piece of the collector 15 16 49 45 46 50 51 17 11 1 Disposable bag with gelling agent The gelling agent influences the extract ed fluid by the amount of powder itself as well as by an additional ap...

Page 16: ... screw 12 2 Foot switch function The foot switch operates pneumatically and has no electrical connection to the VC 45 The unit operating switch 25 for foot pedal operation must be switched to the OFF posi tion The unit is started via an internal switch To switch the appliance ON apply pressure briefly once to the foot switch Another press of the switch will switch the appliance OFF again 30 31 ...

Page 17: ...tion unit fit snugly into the prepared holes of the VC Cart Fix the suc tion unit in place on the VC Cart from below us ing the thumb screws 11 provided 13 2 Moving with the VC Cart Always push the appliance on the cart so that the open fork of the VC Cart is facing in the forward position Ensure the hoses and cables are safe from any possible slipping during transport Never leave the VC Cart on a...

Page 18: ...e the adapter approx 10 cm along the rails provided under the VC Cart and lock in place Place the protective caps 43 over the screws 42 43 25 Slide the collector mount 44 onto the adapt er guide The collector should now hang in the adapter guide Before every application carry out a vis ual check of the complete appliance in cluding the collector If the collector shows any signs of fatigue small cr...

Page 19: ...he elbow piece and T piece Place the connection hose 46 onto the el bow piece 49 and onto the standard adapter piece of the collector 28 Unfold the disposable bag and place inside the collector The elbow piece 50 must now be inserted into the disposable bag Switch on the VC 45 close the connector piece and simultaneously apply slight pressure to the middle of the cover of the disposable bag Briefl...

Page 20: ...rding to the custom ary national regulations for contami nated disposable materials from dental surgeries During disinfection and cleaning it is absolutely vital that the valid acci dent prevention procedures for the surgery are observed e g gloves protective glasses mouth protection Remove the suction handpiece from the suc tion hose and disinfect clean and sterilise ac cording to national rules ...

Page 21: ...in each case The instructions described below have been validated by Dürr Dental as suita ble and are recommended for the prep aration of VC 45 components before and after use The surgery manager is responsible for ensuring that the correct and safe prep aration of the instruments is carried out using suitable appliances materials and trained personnel in order to produce the desired results Valid...

Page 22: ...eria filter excess vac uum stop from the handle section Use both hands to remove the lid from the collector Remove the spray guard Dispose of the contents of the collector as appropriate 16 3 Disinfection and cleaning Use cleaning and disinfection devices strictly according to manufacturer s instructions Place the individual parts into the unit in such a way that the interior surfaces will be rins...

Page 23: ...erilisation with either any other instruments or with the sides of the tray Place parts for sterilisation away from the heating rods and on the middle rack so there is no contact with the inside walls of the system The product must be sterilised in a disassembled state only Temperature 134 C Pressure above the ambient surround ings 2 16 bar 0 216 MPa Working time 5 min Parts for sterilisation must...

Page 24: ...iate Before disposal the parts should be disinfected and sterilised 16 6 Storage Store individual parts protected against contamination 16 7 Disinfection and cleaning of the appliance surfaces Before disinfection and cleaning of the appliance unplug at the mains and re move all power Always clean the surfaces using a damp cloth Do not use a dry cloth In the case of coarse organic contamination pre...

Page 25: ...acteria filter excess suction stop 15 is only to be used once and must be changed at least once a day If the appliance is to be used continu ously for longer periods several hours it is important to change the bacteria fil ter after each patient Never operate the appliance without bacteria filter excess vacuum stop Only use original Dürr bacteria filters Before every application carry out a visual...

Page 26: ...als around collector lid Bacteria filter is blocked display shows vacuum Change bacteria filter check fluid level in the collector if necessary and empty the collector if required Pump valve reeds are stuck Replace pump requires service technician 3 Secretions or blood have been sucked into the pump Working without bacteria filter ex cess vacuum stop Clean the pump of all contamina tion or replace...