Troubleshooting
17 Messages
17.1 "Caution" messages
After acknowledgement the unit will continue to work, but only with limited functions.
Message: Caution
No SD card present. Insert the SD card and restart the device.
SD card write error. Restart the device.
Pressing force too high. Please repeat the sealing process.
Pressing force too low. Please repeat the sealing process.
Sealing temperature too high. Please repeat the sealing process.
Sealing temperature too low. Please repeat the sealing process.
Light barrier triggered for too long.
Sealing speed too high. Please repeat the sealing process.
Sealing speed too low. Please repeat the sealing process.
17.2 "Fault" messages
The unit will no longer function.
When the error has been remedied, it may be necessary to acknowledge the error message.
Message: Fault
Power supply unit defective.
Device overheating. Please wait.
Temperature sensor defective.
Foil transport blocked.
Heating defective. Contact a Service Technician.
Pressing operation defective.
Motor current too high.
17.3 "Note" messages
Important information for the operator, e.g. about the current status of the device.
The unit continues to operate.
Message: Note
Restart device to apply language