16 Cleaning and disinfection
Disconnect all power from the unit.
When cleaning and disinfecting the unit and its
accessories, observe country-specific directives,
standards and specifications for medical prod-
ucts as well as the specific specifications for
dental practices and clinics.
NOTICE
The use of unsuitable agents and
methods can damage the unit and
accessories.
Do not use any products based on phe-
nolic compounds, halogen-releasing
compounds, strong organic acids or
oxygen-releasing compounds, as they
may damage the materials.
❯
Dürr Dental recommends using disin-
fectants from the Dürr Dental product
range. Only the products specified in
these instructions have been subjected
to material compatibility testing by Dürr
Dental.
❯
Read the operating instructions for the
disinfectants.
Wear protective gloves.
16.1 Unit surfaces
The unit surface must be cleaned and disinfected
of any contamination or visible soiling.
Dürr Dental recommends using the disinfectants
FD 322, FD 333 and FD 350.
NOTICE
Liquid can cause damage to the unit.
❯
Do not spray the unit with cleaning and
disinfectant agents.
❯
Make sure that liquid does not get
inside the unit.
❯
Remove any coarse soiling with a soft, lint-free
cloth that has been dampened with cold tap
water.
The unit must be dry before the disinfection
step starts.
❯
Disinfect the surfaces using a disinfection wipe.
Alternatively, use disinfectant on a soft, lint-free
cloth.
Usage
6024100132L02 1906V003
23
EN