Operate the touch screen by tapping it with a
fingertip or the stylus to select a menu or input
field.
Navigating
If the contents of the window cannot be com-
pletely displayed on the touch screen, a scroll bar
appears.
❯
Tap
or
to move the displayed section
of the window.
11.3 Sealing the sterilisation pack-
aging
❯
Place the sterilisation packaging onto the
adjustable stop and guide it into the transport
mechanism until the unit starts up and the ster-
ilization packaging is drawn in (see "13 Adjust-
ment of the adjustable stop").
NOTICE
Incomplete sealing seam or blocked
sterilisation packaging due to wrong
infeed
Incorrect (e.g. half) sealing seam. No/
inadequate foil protrusion for peeling
and labelling (< 1 cm). Foil can get
caught on the transport belt. Foil
remains in the device without an error
message being generated.
❯
Place the sterilisation packaging up
against the stop (this is the only way to
ensure that the light barrier will be trig-
gered).
❯
Observe the maximum foil width of
75 cm.
❯
Assess the sealing seam after process is com-
pleted (see "14.2 Daily function assessment").
❯
Label the packaging with the expiry date and
required additional information (e.
‑
g. ster-
iliser no., personnel, program etc.)
11.4 Start the transport belt manu-
ally
If the sterilisation packaging is not com-
pletely transported out of the unit and
gets stuck, it is possible to manually start
the foil transport (see also"11.2 Touch
screen" ).
❯
On the start screen tap
Foil Transport
.
❯
Press the
button until the sterilisation pack-
aging has been completely transported out of
the unit.
Press and hold the button for continuous
transport.
The transport belt will stop as soon as
you release the button.
❯
Tap
Exit
.
Usage
6024100132L02 1906V003
19
EN