Assembly
6
Requirements
6.1
Installation/setup room
The room chosen for set up must fulfil the follow-
ing requirements:
– Closed, dry, well-ventilated room
– Should not be a room made for another pur-
pose (e. g. boiler room or wet cell)
Ambient and environmental conditions
must be taken into account. Do not oper-
ate the unit in damp or wet conditions.
WARNING
Risk of explosion due to ignition of
combustible materials
❯
Do not operate the unit in any rooms in
which inflammable mixtures may be
present, e.g. in operating theatres.
6.2
Setup options
The following options are available for installation
of the device:
– On a stable, easily accessible surface (e.g.
work table).
Please note the following when setting up the
device:
– The device warms up during operation; keep a
good distance to heat-sensitive devices.
– A minimum gap equal to the width of the
packaging must be provided to the side to
allow insertion and removal of the sterile prod-
uct packaging that is to be sealed.
– Do not cover the ventilation slots.
6.3
Information about electrical
connections
The unit has no main power switch. The unit
must be set up so that the power outlet is easily
accessible and the unit can be unplugged from
the mains if necessary.
7
Installation
7.1
Remove the transport locks
❯
Remove the transport protection.
❯
Store the transport protection
(see also
7.2
Electrical connections
1
2
1
Mains connection
2
Network connection
❯
Before connecting the device, check that the
supply voltage matches the voltage specifica-
tion on the type plate.
❯
Connect the mains cable to the unit and to the
power outlet.
Assembly
12
6024100132L02 1906V003
EN