Attach Scaler handpiece
❯
Attach the Scaler handpiece to the hose con-
nection.
Inserting/changing instruments
CAUTION
If an attempt is made to perform a
treatment with a damaged or worn
instrument, this can cause injury and
may result in an unsuccessful treat-
ment.
❯
Worn or bent instruments should be
replaced immediately and not reused!
The tool kit cover of the instrument is
used as a torque wrench for tightening
the union nut of the instrument chuck.
❯
Select a suitable instrument for the corre-
sponding treatment.
To prevent the instrument from being
overtightened, the torque wrench slips
when the required torque is reached. No
ratcheting sound can be heard.
❯
Always use the torque wrench to screw the
instrument on and off.
When screwing the instrument on, slowly turn
the torque wrench approx. a quarter turn
beyond the resistance point.