Durham 9840 Setting Up Download Page 1

Setting up the 

9840 Load Cell Indicator

To turn on the 9840 Load Cell 

Connector, reach behind in the back 

corner and flip the switch up.

Before starting anything, be sure to first 

plug in the sensor(s) you will be using.

In starting out, you will see 

“>User Data Entry” on the screen.

To enter setup mode, press the two end 

buttons at the same time  (

 and 

).

(1)

(2)

(3)

As it starts up, you will see a series of 

screens introducing you to the 9840 Load 

Cell Connector and displays of the 

settings from its last use.

(4)

7418 East Helm Drive • Scottsdale, Arizona 85260 • 480.948.5555 • www.interfaceforce.com

 Page 1 of 

7

Setup Guide

 9840  v

1.1

  0

9

-

18

-2018

Summary of Contents for 9840

Page 1: ... using In starting out you will see User Data Entry on the screen To enter setup mode press the two end buttons at the same time and 1 2 3 As it starts up you will see a series of screens introducing you to the 9840 Load Cell Connector and displays of the settings from its last use 4 7418 East Helm Drive Scottsdale Arizona 85260 480 948 5555 www interfaceforce com Page 1 of 7 Setup Guide 9840 v1 1...

Page 2: ...ER After being brought to the next screen press till you get to the Calibration option and then press ENTER Before you can enter the calibration settings you will be required to enter a password 7418 East Helm Drive Scottsdale Arizona 85260 480 948 5555 www interfaceforce com Setup Guide 9840 v1 1 09 18 2018 Page 2 of 7 ...

Page 3: ... the measurement cell type that you are using For this example we are using a load cell so we select Load Then press ENTER 10 Enter the cell s serial number by using the or signs then for the next digit ENTER 7418 East Helm Drive Scottsdale Arizona 85260 480 948 5555 www interfaceforce com Setup Guide 9840 v1 1 09 18 2018 Page 3 of 7 ...

Page 4: ...git s Do the same for the day and year and press ENTER after each one Select the necessary unit of measure and press ENTER 11 12 7418 East Helm Drive Scottsdale Arizona 85260 480 948 5555 www interfaceforce com Setup Guide 9840 v1 1 09 18 2018 Page 4 of 7 ...

Page 5: ...r this example the max capacity of the 1210AF is 10K so we enter 10000 and then press ENTER Or you can enter the appropriate SEB Output Value as your full scale output value 13 Enter the voltage excitation using buttons and T hen press ENTER 14 7418 East Helm Drive Scottsdale Arizona 85260 480 948 5555 www interfaceforce com Setup Guide 9840 v1 1 09 18 2018 Page 5 of 7 ...

Page 6: ...ttons and Then press ENTER For the next display you will be required to enter the mV V for the full scale output of the direction that you are going to use In this example we will pick Tension We see that the value of 4 07906 is the full scale mV V correlating with the max capacity of 10K that we entered earlier 7418 East Helm Drive Scottsdale Arizona 85260 480 948 5555 www interfaceforce com Setu...

Page 7: ...8a 18b Calibration setup is now complete To return to the main menu press ESC 7418 East Helm Drive Scottsdale Arizona 85260 480 948 5555 www interfaceforce com Setup Guide 9840 v1 1 09 18 2018 Page 7 of 7 ...

Reviews: