Instructions for Use

TRI-CLEAN

®

BrandMax

®

www.

DuralineSystems

.

com

1. Fill Unit

n

   Move drain valve to closed position.

n

   At the beginning of each day you will need to “Degas” your unit. In order to perform this

important function please fill you ultrasonic cleaner to the max fill line and run the unit for 
15 minutes. This will minimize the microscopic bubbles in the tank and improve cleaning 
effectiveness.

n

   If heat is desired please adjust to appropriate temperature (110

o

F) using the arrows labeled 

“Temp” then press the “Heating” button to initiate the feature. (Note: Actual temperature will 
display once heat is started and may take up to 30 minutes to reach desired temperature).

n

   After completing the Degas operation, please add the correct amount of enzymatic cleaning 

solution.

 (Tri-Clean triple enzymatic cleaner = 1 oz. per gal.)

n

   Carefully place items to be cleaned in basket and lower into solution.

n

   Place lid on the top of the unit.

2. Start Cleaning Cycle

n

   Caution, while using heat does improve your cleaning results, it also results in the solution

becoming hot. Please use extreme caution when operating the unit and do not place hands in 
the solution or spill solution on you. 

n

   Set desired time of operation by using the arrows labeled “Time”. 

n

   Start the cleaning cycle by pressing the button labeled “Ultrasonic”.

3. Unit Operation

n

   Wait for cleaning cycle to finish or terminate the cycle early by pressing the ultrasonic button.

(Note: Do not stop the cycle by using the large green illuminated switch labeled “Power”).

n

   Turn off power to ultrasonic cleaner by using the green illuminated switch labeled “Power”. 

(Note: Power switch for recessed models is not illuminated and located on side of control panel).

n

   Carefully remove basket from solution and allow excess liquid to drain off before moving away 

from tank. 

n

   Drain contents of tank by opening the valve located on the side of unit. (Note: Recessed models 

must be powered on to activate drain valve by pressing button labeled “Drain” on control panel).

n

   Flush tank with warm water and clean with a non-acetone or non-phenol based cleaner if film is 

present on tank walls.

Please visit 

www.bit.ly/cleaninstrument

 

for basic tips on effective instrument cleaning. 

Note – please contact the instrument manufacturer to confirm cleaning guidelines.