![DURAG D-R 290 Operating Manual Download Page 85](http://html.mh-extra.com/html/durag/d-r-290/d-r-290_operating-manual_2541137085.webp)
5 | Installation
D-R 290
85
1
Instructions for aligning the measuring head/reflector at the adjustment flange:
First tighten nut B.
Vertical adjustment: with nut C Horizontal adjustment: with nut A
2
Duct wall
3 Welded-in pipe with adjustment
flange
4
Red dot
5 Threaded studs
6
Rubber sleeve
7 12 disc springs *
8
1 spherical washer *
9 1 nut *
10 measuring head (M) or reflector (R)
11 Purge air flange
12 Drive
13 Fail-safe shutter
14 Flat gasket
* per threaded stud
Fig. 5.18: Installation at the adjustment flange, dimensional diagram D−R 290
When installing the measuring head and reflector with protective devices (fail-safe
shutters) on the adjustment flange, perform the operations in the following sequence:
1.
- 14) on to the threaded studs of the welded-in
flanged tube.
2.
Also slide the D−SK 290 MA fail-safe shutter on to the threaded studs. When doing
so, align the motor housing to the side facing away from the duct and facing down
wards (
- 12).
3.
First pull the rubber sleeve (
- 6) over the purge air flange (11) of the
measuring head or reflector.
Summary of Contents for D-R 290
Page 12: ...12 D R 290 ...
Page 18: ...1 General 18 D R 290 ...
Page 20: ...20 D R 290 ...
Page 28: ...2 Safety 28 D R 290 ...
Page 30: ...30 D R 290 ...
Page 38: ...3 Delivery 38 D R 290 ...
Page 40: ...40 D R 290 ...
Page 60: ...4 Product description 60 D R 290 ...
Page 106: ...106 D R 290 ...
Page 152: ...152 D R 290 ...
Page 174: ...8 Maintenance 174 D R 290 ...
Page 176: ...176 D R 290 ...
Page 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Page 194: ...9 Appendix 194 D R 290 ...
Page 200: ...11 Index 200 D R 290 ...
Page 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...