5 | Installation
78
D-R 290
1
Instructions for aligning the measuring head/reflector at the adjustment flange:
First tighten nut B.
Vertical adjustment: with nut C Horizontal adjustment: with nut A
2
Duct wall
3 Welded-in pipe with adjustment
flange
4
Red dot
5 Threaded studs
6
Rubber sleeve
7 12 disc springs *
8
1 spherical washer *
9 1 nut *
10 measuring head (M) or reflector (R)
11 Purge air flange
* per threaded stud
Fig. 5.10: Installation at the adjustment flange, dimensional diagram D−R 290
When installing the measuring head and reflector without protective devices (fail-safe
shutters) at the adjustment flange, perform the operations in the following sequence:
1.
First pull the rubber sleeve (
– 6) over the flange tube in the measurement
duct (3)
2.
Connect the purge air hose to the purge air adapter of the measuring head or re
flector (11) and secure it with the hose clamps supplied.
Summary of Contents for D-R 290
Page 12: ...12 D R 290 ...
Page 18: ...1 General 18 D R 290 ...
Page 20: ...20 D R 290 ...
Page 28: ...2 Safety 28 D R 290 ...
Page 30: ...30 D R 290 ...
Page 38: ...3 Delivery 38 D R 290 ...
Page 40: ...40 D R 290 ...
Page 60: ...4 Product description 60 D R 290 ...
Page 106: ...106 D R 290 ...
Page 152: ...152 D R 290 ...
Page 174: ...8 Maintenance 174 D R 290 ...
Page 176: ...176 D R 290 ...
Page 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Page 194: ...9 Appendix 194 D R 290 ...
Page 200: ...11 Index 200 D R 290 ...
Page 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...