4 | Product description
D-R 290
47
Op
Opacity at the measuring point (value corresponding to a single pas
sage)
Op1
(measured) opacity at the measuring point
(value corresponding to a double passage;
Op2
Opacity at the flue stack opening
(value corresponding to a single passage)
L1
Diameter of the flue stack at the measuring point
L2
Diameter at the flue stack opening
PLCF
Path length correction factor (flue stack correction factor)
Table 4.4: Legends in the formulae
Table 4.5: Flue stack correction factor
Equation
4
Table 4.6: Equation 4
Based on the (measured) double attenuation of the measurement light at the measur
ing point, we have the following relationship:
Equation
5
Table 4.7: Equation 5
Flue stack mouth
Measuring point
Fig. 4.9: Reference points for determination of the opacity