![DURAG D-R 290 Operating Manual Download Page 155](http://html.mh-extra.com/html/durag/d-r-290/d-r-290_operating-manual_2541137155.webp)
8 | Maintenance
D-R 290
155
●
the general ambient circumstances (e.g. climatic conditions at the measuring point)
It makes sense to start with a short maintenance interval (typically 4 weeks). The
maintenance interval can then be increased step by step up to a maximum of 6
months, in accordance with the prevailing conditions.
Maintenance should be performed during an inspection of the system being moni
tored, i.e. this means the system is not in operation.
Irrespective of this however, a check should be performed every four weeks. The
check is to look for external changes such as deposits, faulty seals, colour changes on
the casing, etc. It is also a check that all screw connections, closures and devices are
tightly secured.
If the measuring system is not maintained or insufficiently maintained, increasing con
tamination of the lens and filter, if necessary in accordance with the respective instruc
tions for the system, may lead to the failure of the measuring head and possible dam
age to it.
Table of maintenance work (suggestion)
K* Activity
Who
When
a
Visual inspection and cleaning of external parts of the device
Operating personnel** every four weeks
a
Checking the closures and screw connections
Operating personnel** every four weeks
a
Check the purge air unit and the hose connections for leaks
and secure fitting
Operating personnel** every four weeks
b
Remove deposits on and within the welded-in pipe
Operating personnel** every four weeks
b
Purge air blower:
Checking the filter
Changing the filter
(see also Checking / changing the filter)
Specialised staff**
every four weeks
typically every
3 months
b
Cleaning the purge air hoses
Specialised staff**
Typically every
3 months
b
Measuring head:
Checking the contamination value (see --- FEHLENDER LINK
---),if necessary cleaning the device glasses
Specialised staff**
every four weeks
b
Check the performance LED alignment
Specialised staff**
Typically every
3 months
a
Check the measured values for contamination, zero point and
reference point
Operating personnel** every four weeks
b
Linearity check (see Checking the linearity)
Specialised staff**
1/year
* K ≘ Category - ** Definition see section
Table 8.1: Suggestion for checking and maintenance work
Category for maintenance work
Work of category "a" can be performed whilst the plant is in operation (the work in
structions are created by the operating company). Work of category "b" requires the
purge air to be switched off.
When performing any maintenance work of category "b", ensure that:
The purge air is necessary for protection of the component units. The purge air shields
components from hot and/or aggressive gases. Loss of the purge air supply even for a
(very) short time can lead to malfunctions, damage or total write-off of the measuring
head.
Summary of Contents for D-R 290
Page 12: ...12 D R 290 ...
Page 18: ...1 General 18 D R 290 ...
Page 20: ...20 D R 290 ...
Page 28: ...2 Safety 28 D R 290 ...
Page 30: ...30 D R 290 ...
Page 38: ...3 Delivery 38 D R 290 ...
Page 40: ...40 D R 290 ...
Page 60: ...4 Product description 60 D R 290 ...
Page 106: ...106 D R 290 ...
Page 152: ...152 D R 290 ...
Page 174: ...8 Maintenance 174 D R 290 ...
Page 176: ...176 D R 290 ...
Page 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Page 194: ...9 Appendix 194 D R 290 ...
Page 200: ...11 Index 200 D R 290 ...
Page 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...