7 | Checking / setting parameters with D‑ESI 100
146
D-R 290
After the blockage has been removed, "0" must by input here to allow the shutter to be
opened again.
Every time the device is restarted the parameter "
Fail-safe shutter 1 (or 2) blocked and
deactivated" is set to "0". The system then attempts to open the shutter(s).
Calibration on the dust-free measuring path
The calibration is performed in the following sequence:
1.
Install the measuring head and reflector in a dust free space (measuring tube) at
the exact original measuring path length. The lengthening of the path due to the
disc springs and the fail-safe shutters must also be taken into account. Ensure the
most parallel alignment possible of the flanges for the measuring head and the re
flector.
2.
The optical surfaces (outer glass and zero point reflector) of the measuring system
must be cleaned very carefully and without leaving any smears, using an optical
cleaning cloth and glass cleaning agent (see section
8.2.4 Cleaning the outer glass
and the zero-point reflector: [
).
3.
Switch on the measuring head and wait until it is ready for operation (the mainte
nance LED must not light up continuously).
4.
Check the contamination display. The contamination must not be greater than 2%.
Setting path:
Device (D‑R 290 ) > >
Specific Parameters >
D‑R 290
Measured values > contamination [%]
1.
First the measuring head has to be aligned optically (see section
). If the measuring head and the reflector have been mounted as
or
, the 3 nuts on the reflector flange can be
tightened.
First tighten the nut B on the measuring head. When the nut A is tightened the
measuring head inclines to the axis B - C, and the figure in the image display
moves in a vertical direction. When the nut C is tightened or loosened, the measur
ing head pivots at the axis A - B and the figure in the image display moves in a
horizontal direction.
2.
The metal knurled screw on the optics has to be removed (control the focus set
ting) and the focus has to be set with the plastic button so that a sharp image can
be seen in the image display for control distances of 1 to 2.25 meters (reflector 1).
For control distances of more than 2.25 metres (reflector 2), turn the plastic button
anticlockwise to the stop.
3.
Then lock the setting with the metal knurled screw.
4.
The measurement on the dust-free measuring path can be started in the area
Functions -> D‑R 290 Service Functions (see illustration below).
Setting path:
Device (D‑R 290 ) > >
Specific Parameter >
D‑R 290
Measured values > Contamination [%]
7.7.8
Step 1
Step 2
Summary of Contents for D-R 290
Page 12: ...12 D R 290 ...
Page 18: ...1 General 18 D R 290 ...
Page 20: ...20 D R 290 ...
Page 28: ...2 Safety 28 D R 290 ...
Page 30: ...30 D R 290 ...
Page 38: ...3 Delivery 38 D R 290 ...
Page 40: ...40 D R 290 ...
Page 60: ...4 Product description 60 D R 290 ...
Page 106: ...106 D R 290 ...
Page 152: ...152 D R 290 ...
Page 174: ...8 Maintenance 174 D R 290 ...
Page 176: ...176 D R 290 ...
Page 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Page 194: ...9 Appendix 194 D R 290 ...
Page 200: ...11 Index 200 D R 290 ...
Page 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...