6 | Commissioning
108
D-R 290
2.
Once the measuring head and reflector have been installed as shown in
or
, the 3 nuts on the respective flanges are tightened as described in the
points below.
3.
By this means the optical alignment of the measuring head is performed at the
same time.
For this a sighting mechanism (Fig. on the left middle) is available on the measur
ing head, which is located on the right side behind a sight glass. The image of the
reflector (spot of light) must be aligned so that it is positioned in the centre of the
sighting mechanism (see figure on the left, bottom). If necessary the reflector must
also be aligned (at right angles to the gas duct).
4.
Now nut B is tightened first (Fig. on the left, top).
When the nut A is tightened, the measuring head changes its inclination relative to
the axis B - C and the image in the sighting mechanism moves in a horizontal di
rection. Tightening or releasing the nut C causes a pivoting movement about the
axis A - B and the image in the sighting mechanism moves in the vertical direction.
Only tighten the (self-locking) nuts to such a degree that the disc springs can still
perform their spring action.
Parameterisation and operation of the measuring head
The device is supplied with factory settings. The calibration to the dust-free measuring
path is performed in the factory provided that the operating company has supplied a
completed measuring point questionnaire. After successful initial commissioning, the
parameterisation of the measuring head must be performed by software. The basic
parameters necessary for operation must be saved.
DURAG service technicians will be glad to assist you if required. They have the nec
essary equipment and software. You can find the addresses and phone numbers of
the service engineers under DURAG GROUP company addresses.
Specific hardware and software are required for inputting or checking parameters. We
recommend the use of the optional D‑ESI 100 SET.
The standard interface for parameterisation is the USB port. To allow the use of this
port, the protective cap must be opened on the USB port of the measuring head (see
Fig. 6.1: Sighting mechanism
Fig. 6.2: Optical alignment
6.2
Summary of Contents for D-R 290
Page 12: ...12 D R 290 ...
Page 18: ...1 General 18 D R 290 ...
Page 20: ...20 D R 290 ...
Page 28: ...2 Safety 28 D R 290 ...
Page 30: ...30 D R 290 ...
Page 38: ...3 Delivery 38 D R 290 ...
Page 40: ...40 D R 290 ...
Page 60: ...4 Product description 60 D R 290 ...
Page 106: ...106 D R 290 ...
Page 152: ...152 D R 290 ...
Page 174: ...8 Maintenance 174 D R 290 ...
Page 176: ...176 D R 290 ...
Page 178: ...9 Appendix 178 D R 290 Fig 9 2 Measurement point questionnaire 2 ...
Page 194: ...9 Appendix 194 D R 290 ...
Page 200: ...11 Index 200 D R 290 ...
Page 202: ...GmbH Kollaustraße 105 22453 Hamburg Germany www durag com ...