User Training 

 
The end user of the VC225ILR refrigerator should be instructed in the simple processes outlined 
below and also those in 

“Section 11 Maintenance tasks”. It should be stressed that these tasks are 

essential to ensure correct and reliable operation of the refrigerator. 
 

Caution:  failure  to  follow  these  instructions  may  lead  to  system  failure  and  vaccine 
wastage. 

 

If you have a problem, please contact your appointed engineer. 

 

Turning on 

 
When  turning  on  for  the  first  time  or  when  the  refrigerator  has  been  turned  off  for  more  than  3 
days: 
 

  Turn the power on and allow the refrigerator to run in accordance with the table below: 

 

Total No. of hours the power is off 

each day 

Running time required before full 

holdover is achieved 

(at +43

°

C ambient) 

Continuous operation possible 

once cool down has been 

achieved 

2 days 

12 

4.5 days 

16 

7 days 

20 

14 days 

 

  Before loading with vaccines, ensure that the temperature display on the front of the 

refrigerator reads +4.5°C.  

 

Daily Routine 

 

  Keep the lid closed, open only when required and lock after use. 

  Store the most frequently used vaccines in the top baskets for easy access. 

  Do not keep food or drink in the refrigerator. 

 

Vaccine Storage: 

 

  Only store vaccines in the refrigerator. 

  Always store vaccine in its original packaging. 

  Always keep the same type of vaccine together and store your vaccine neatly. 

  Always use old vaccines before new ones; before use make sure the vaccine has not 

expired. 

Summary of Contents for VC225ILR

Page 1: ...DULAS VC225ILR Mains Powered Vaccine Refrigerator User Manual Ref MU VC225ILR ENG F ...

Page 2: ...STRUCTIONS 4 3 YOUR DULAS MAINS POWERED REFRIGERATOR 6 4 REFRIGERATOR COMPONENTS 7 5 MATERIAL SAFETY INFORMATION 8 6 END OF LIFE PRODUCT RECYCLING 9 7 REFRIGERATOR INSTALLATION 11 8 REFRIGERATOR LOADING 12 9 STATUS LIGHT 13 10 USER TRAINING 14 11 MAINTENANCE TASKS 15 12 VC225ILR SYSTEM LAYOUT 15 ...

Page 3: ... QUANTITY REFRIGERATOR VC225ILR Refrigerator Cabinet 1 Refrigerator Baskets 6 Refrigerator Key 2 Other Power Supply PS ILR 001 1 Screws 5x40mm for wall mount 5 Wall plugs Brown for wall mount 5 Screws 10 Wall plugs Red 10 Cable ties 10 Cable tie bases 10 MC connection tool 1 ...

Page 4: ... must be supervised not to play with the appliance This refrigerator is for professional use only It is not designed for household use Refrigerator and power system are heavy please observe good manual handling procedures when moving the refrigerator and solar power system Only use suitable transport equipment Live electrical components The refrigerator operates from low voltage DC There are high ...

Page 5: ...refrigerant Due to this fact the following information is particularly important Warning Do not damage the refrigeration circuit Ensure that sharp or pointed objects do not come into contact with the refrigeration circuit Warning Do not use mechanical devices or other means to accelerate the defrosting process other than those recommended by the manufacturer Warning Keep ventilation openings in th...

Page 6: ...erature when the power is off the Dulas ILR has a phase change lining This lining acts as a cold store and stores energy that keeps the temperature inside the refrigerator stable when electricity is not available The temperature of the cold store in the Dulas ILR is naturally between 4 to 5 C which is the perfect temperature required for safe vaccine storage and means that the product is 100 Freez...

Page 7: ...Dulas VC225ILR User Manual Page 7 of 15 4 Refrigerator Components ...

Page 8: ...red inside the walls of the refrigerator In normal use the user will never be aware of its presence This PCM is classified under regulation EC No 1272 2008 as a category 1 aspiration hazard and is classified under regulation 67 548 EWG 1999 45 EG as having the potential to cause physical harm R65 R66 At normal ambient temperatures this product will be unlikely to present an inhalation hazard becau...

Page 9: ...igerator contain an organic phase change material that can be harmful and must be disposed of safely Please contact Dulas for advice Worn out devices must be professionally disposed of in accordance with local regulations and legislation Potentially hazardous components materials are Refrigerant gas R600a Phase change material PCM paraffins normal C5 20 Electrolytic capacitor Electronic components...

Page 10: ...nd its implementation as national law electrical equipment that has reached the end of its life must be collected separately and returned to an approved recycling facility Any device that you no longer require must be returned to our agent or find out about the approved collection and recycling facilities in your area Ignoring this European Directive may have potentially adverse effects on the env...

Page 11: ...fficient slack in the cable to allow access to the back of the refrigerator 4 Plug the Dulas PS ILR 001 into a suitable AC outlet and turn it on Then switch the DC isolator to the 1 position Within 1 minute the compressor will start if there is AC power You will be able to hear the compressor operating and feel the condensers getting warm 5 Check that the refrigerator compartment temperature displ...

Page 12: ...e vaccine compartment that drop below freezing and thus no danger of freezing vaccines Optional stacking baskets are available These are designed to make stock management easier and encourage air circulation within the compartment Try to keep the lid open for as short a time as possible Stock management Allocate a certain space for each different type of vaccine so that they may be located rapidly...

Page 13: ...wer Normal operation The refrigerator warms up slightly overnight It is usual for the compressor to start running shortly after sunrise It will run continuously for a few hours to cool the refrigerator After this time the compressor will begin cycling on and off to maintain the correct internal temperature This will continue until the sun sets at the end of the day At night there is no power to th...

Page 14: ... in accordance with the table below Total No of hours the power is off each day Running time required before full holdover is achieved at 43 C ambient Continuous operation possible once cool down has been achieved 0 2 days Y 12 4 5 days Y 16 7 days Y 20 14 days Y Before loading with vaccines ensure that the temperature display on the front of the refrigerator reads 4 5 C Daily Routine Keep the lid...

Page 15: ...s Every morning and afternoon Check the temperature Fill in the daily record sheet On the first day of every month Clean the refrigerator condensers and compressors Every 6 months Check all mechanical fixings and electrical connections 12 VC225ILR System Layout ...

Reviews: