TECHNICAL NOTE 

 

392-0027-029 Rev 61 

Configuring the F40-TT transmitter in Dataquest A.R.T. 

The F40-TT was formally added as a transmitter model in Dataquest A.R.T. v4.1; however, 
Dataquest A.R.T. v4.0 and earlier versions can be used with a custom configuration.  
Additionally, as of June 2008, the F40-TT is being calibrated over an extended temperature 
range (22-41ºC) to provide accurate measurements of tail temperature.  This document 
outlines the steps used to configure the F40-TT transmitter including the adjustments in the 
software that are required to accommodate the extended temperature range. 

Important!  The following steps are required in A.R.T. version 4.1 and earlier to ensure 
the accuracy of temperature measurements even if the F40-TT device is not being 
used over the extended temperature range

.  This document is a supplement to the 

configuration information found in the Dataquest A.R.T. user manual. 

No changes in the software configuration are required for A.R.T. v4.2. 

Dataquest ART version 4.1 

1.  In Dataquest A.R.T. Acquisition, go to 

Hardware|Configuration 

 

2.  Highlight the receiver in which the F40-TT will be used, right-

click and choose “

New 

Transmitter

”.

 

 

3. 

Select the model name ‘

TL11M3-F40-TT

’ from the list of transmitters and enter the 

device’s unique serial number and select Next>

 

 

 

 

4.  Select the species and create a unique Animal ID name, select Finish>

 

 

Summary of Contents for F40-TT

Page 1: ...perature range Important The following steps are required in A R T version 4 1 and earlier to ensure the accuracy of temperature measurements even if the F40 TT device is not being used over the exten...

Page 2: ...ing so go to the Advanced Tab and adjust the temperature values to 23 31 39 as shown 6 Optionally the calibration values can be entered here or you may revert back to the Calibrations tab 7 Enter the...

Page 3: ...battery voltage levels and calibrations found on the sterile package and select OK New calibration temperatures from the previous step appear here Note The calibration values shown are examples Use t...

Page 4: ...A R T version 4 0 and earlier 1 In Dataquest A R T Acquisition go to Hardware Configuration 2 Highlight the receiver in which the F40 TT will be used right click and choose New Transmitter 3 Select t...

Page 5: ...al ID name select Finish 5 A new transmitter will appear in the configuration window with 4 channels listed as unconverted 6 Select and right click the first unconverted signal and choose Properties 7...

Page 6: ...8 Go to the Advanced Tab and adjust the temperature values to 23 31 39 as shown...

Page 7: ...9 Optionally the calibration values can be entered here or you may revert back to the Calibrations tab 10 Enter the calibration values found on the sterile package and select OK Critical step...

Page 8: ...nd on the sterile package and select OK Please note that for A R T versions 2 2 and earlier the voltage signal type is not available Channel 3 should be left as unconverted and the battery voltage dat...

Page 9: ...ted designation 15 The device is now configured and should appear as shown Repeat steps for additional transmitters or exit Hardware Configuration 16 In the main Acquisition window double click the an...

Page 10: ...Source Settings Remove the check for the Period data in waveforms and parameters so it is not saved online It is not necessary to save these data uncheck uncheck...

Reviews: