Drucker Diagnostics Horizon 6 Flex Operator'S Manual Download Page 1

HORIZON                

6 Flex FA 

Operator’s Manual

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Summary of Contents for Horizon 6 Flex

Page 1: ...HORIZON 6 Flex FA Operator s Manual...

Page 2: ...atements 3 Initial Setup 4 Operation 4 Quick Start 5 Single Cycle Lock 6 Preset Lock Error Bookmark not defined Setting or Modifying a Cycle 6 Settings 7 Balancing Loads 7 Care and Preventative Mainte...

Page 3: ...Electrical and electronic products recycling symbol Recycle only as electronic waste Do not dispose in normal waste RoHS Compliant Compliance with RoHS environmental standards CE Mark Denotes conform...

Page 4: ...rers IFUs o Lid lighting indicates the centrifuge s status ready running done informing the operator when tubes are ready for the analyzer and preventing tubes from being left in the centrifuge longer...

Page 5: ...materials of a higher risk group are being used more than one level of protection must be provided The use of flammable or explosive materials as well as those materials which have a vigorous chemical...

Page 6: ...Plug the line cord into the centrifuge o Plug the line cord into an approved electrical outlet o Turn on the power switch in the back of the centrifuge BE SURE THE ELECTRICAL OUTLET IS ALWAYS ACCESSIB...

Page 7: ...lock Allows access into the rotor chamber by engaging the unlocking mechanism Entry is only possible when the rotor is stopped Stop Pressing the UNLOCK button during operation will terminate the run a...

Page 8: ...ters Use the left and right arrow button until the parameter to be edited is blinking and underlined Change Parameter Values To change the parameter shown on the display use the up and down buttons ne...

Page 9: ...Spinning balanced loads will extend the life of the centrifuge and produce better results Use the following rules when loading the rotor If an odd number of samples is to be spun fill a tube with wate...

Page 10: ...rtingor storing the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there i...

Page 11: ...them o Inspect the rotor tube holders or buckets for damage o If the tube holders or buckets have any damage even slight safely dispose of them and replace them o If the rotor appears damaged contact...

Page 12: ...nlock by conventional means Remove the latch label and use a pen to manually disengage the locking mechanism Pull the mechanism towards the control panel and then unlatch and open the lid o If the uni...

Page 13: ...ge 16 32 degree C Voltage 115 230 VAC Frequency 50 60 Hz Power Requirement 200 Watts Centrifuge Motor H P Brushless Max g Force 1 600 xg Max Speed 3 500 RPM 100 Cycle Time 1 to 30 minutes 2 CALCULATIN...

Page 14: ...ng of waste electrical and electronic equipment The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recy...