Drucker Diagnostics DASH COAG Operator'S Manual Download Page 8

Drucker Diagnostics – Customer Service: +1-814-692-7761 – 

[email protected]

 | Page 2 

W

ARRANTY

 

Drucker Diagnostics warrants that this centrifuge is free from defects in workmanship and parts for 2 years.

 

 

WARNING: For the safety of both the operator and service personnel, care should be taken when using 
this centrifuge if handling substances that are known to be toxic, radioactive or  contaminated with 
pathogenic microorganisms. Use appropriate personal protection equipment (PPE). When Risk Group II 
materials are used, (as identified in the World Health Organization “Laboratory Bio-Safety Manual”), a Bio- 
Seal should be employed. In the event that materials of a higher risk group are being used, more than one 
level of protection must be provided. The use of flammable or explosive materials as well as those 
materials which have a vigorous chemical reaction is prohibited. 

 

Unplug the centrifuge before cleaning or performing maintenance. 

 

WARNING: Inspect centrifuge for cracks or physical damage to cabinet, lid, rotor, or tube holders. Damage 
may result in unsafe operation. Discontinue use until repairs have been performed. 

 

This equipment generates, uses, and can radiate radio frequency energy, and if not installed and used in 
accordance with this operator manual, may cause interference to radio communications. 

 

Operation of this equipment in a residential area may cause interference, in which case the user will be 
required to correct the interference at his own expense.  

 

I

NITIAL 

S

ETUP 

 

o

 

Unpack and verify that all the following are included: 

 

 Centrifuge 

 

 Power cord 

 

 Tube holders 

 

 Operating manual 

o

 

Setup the centrifuge on flat and level surface. A bench top clearance height of 21” (54 cm) is required to open 

the lid. 

o

 

The centrifuge should have 6” (15 cm) of clear space around the centrifuge. Proper ventilation is necessary to 

prevent the overheating of samples as well as premature failure of the centrifuge. Choose an area which allows 

unencumbered air flow, and where the temperature remains between 16°C and 32°C. 

o

 

No hazardous material shall be permitted in the clearance envelope during operation.  

o

 

The operator time within the envelope shall be limited to the time necessary for loading, unloading,  and 

centrifuge operation only. 

o

 

Plug the line cord into an approved electrical outlet. 

 

B

E SURE THE ELECTRICAL OUTLET IS ALWAYS ACCESSIBLE AS THE LINE CORD IS THE MEANS OF EMERGENCY 

D

ISCONNECTION

 

 

Summary of Contents for DASH COAG

Page 1: ...DASH COAG Operator s Manual...

Page 2: ......

Page 3: ...ial Setup 2 Quick Start 3 Settings 4 Review Cycle Time and Speed Settings 4 Changing Cycle Time and Speed Settings 4 Operation 5 Balancing Loads 5 Care and Preventative Maintenance 6 Cleaning and Disi...

Page 4: ......

Page 5: ......

Page 6: ......

Page 7: ...to reinforce the tube holders and provide high strength and durability o A clear lid permits safe observation of samples and optical calibration of speed o The lid safety system only allows entry into...

Page 8: ...repairs have been performed This equipment generates uses and can radiate radio frequency energy and if not installed and used in accordance with this operator manual may cause interference to radio c...

Page 9: ...ng the locking mechanism Entry is only possible when the rotor is stopped Stop Pressing the UNLOCK button during operation will terminate the run and unlock the lid after the rotor has come to astop C...

Page 10: ...in the current cycle 10 beeps flashes equal 10 minutes of run time Run time starts when the rotor reaches 90 of desired speed and stops when the rotor starts decelerating o Pressing the START button a...

Page 11: ...will slow to a complete stop and the lid light will flash o The locking mechanism will disengage for 60 seconds allowing entry into the rotor chamber To unlock after more than 60 seconds have elapsed...

Page 12: ...tingor storing the centrifuge to prevent damage andinjury CLEANING AND DISINFECTION To prolong the life of the centrifuge cleaning and disinfection is recommended every six months or whenever there is...

Page 13: ...mage o If the tube holders or buckets have any damage even slight safely dispose of them and replace them o If the rotor appears damaged contact Customer Service for further assistance Excessive noise...

Page 14: ...o change the run preset time The lid does not open o Ensure that the lid knob is turned fully counterclockwise o If the knob cannot be turned counterclockwise turn it fully clockwise press UNLOCK and...

Page 15: ...0 Watts Centrifuge Motor H P Brushless DC Max G Force 4 400 xg Max Speed 6 600 RPM Min Cycle Time 1 minute Max Cycle Time 30 minutes Use only with approved accessories from the original manufacturer A...

Page 16: ...ectronic equipment The separate collection and recycling of your waste equipment at the time of disposal will help to conserve natural resources and ensure that it is recycled in a manner that protect...