Copyright 2010 Baker Hughes Company.
English–PACE Indicators Instruction Manual | 29
Introduction
4. Maintenance
4.1 Introduction
This section contains procedures for routine maintenance and the replacement of components.
Refer to Section
5, “Testing and Fault Finding,” on page
31.
4.2 Visual Inspection
Inspect for obvious signs of damage and dirt on the following:
a. External of the instrument.
b. Power supply adaptor.
c.
Associated equipment.
Damaged parts must be replaced, contact Druck Service.
4.3 Cleaning
Do not use solvents for cleaning. Clean the front panel with a damp lint-free cloth and mild
detergent.
4.4 Test
Do a standard serviceability test. Refer to Section
5.2, “Standard Serviceability Test,” on
4.5 Software Update
To update the PACE1000 internal software, use the following procedure:
1. Insert a USB memory device into a web connected PC.
2. Open Windows Explorer and select the USB memory device root folder. Delete the following
folders if they are present:
i.
DPI
ii.
OS
3. Using a web browser go to the following Druck PACE support page:
https://www.bakerhughesds.com/download-center?bh_ref=gemeasurement.com
Table
4
-
1: Maintenance Tasks
Task
Period
Visual Inspection
Before use
Test
Before use
Cleaning
Weekly
a
a. May change; depends on usage (e.g. rack mounted,
bench top) and environment (e.g. humidity, dust).
Calibration
12 months
b
b. May change; depends on the required accuracy.
Summary of Contents for PACE1000
Page 1: ...Druck com PACE Indicators Pressure Automated Calibration Equipment Instruction Manual...
Page 2: ......
Page 10: ...Copyright 2010 Baker Hughes Company viii PACE Indicators Instruction Manual English...
Page 29: ...Copyright 2010 Baker Hughes Company English PACE Indicators Instruction Manual 19 Data Logging...
Page 75: ......