KA215 Issue No. 2

Druck Incorporated, 4 Dunham Drive, New Fairfield, Connecticut 06812, U.S.A.

Tel: (203) 746 0400

Fax: (203) 746 2494

Email: [email protected]

Internet  http://www.druckinc.com

Preparation

1.
Connect the instrument
to a pressure source
that has an accuracy
three times better than
the instrument.
Recommended:-
Druck DPI 610 or DPI
610 IS Portable
Calibrator.

2.
Switch on the instrument
and select the units of
pressure measurement
required for calibration.

Procedure

1.
Press all three keys
together to enter the
CAL menu and proceed
as shown:

Changing the PIN

Each digit can be
changed in turn.
Pressing:

increases the
value of the digit.

steps to next digit,
left
to right.

enters the PIN

Druck DPI 705 and DPI 705 IS
Digital Pressure Indicators

Calibration Instructions

WARNING: CALIBRATE DPI 705 IS INSTRUMENTS IN A SAFE AREA.

a

d

The instrument performs a two-point compensation at ZErO and FS (full-scale.)