background image

Med-Aire Edge  

Alternating Pressure & Low Air Loss

Mattress Replacement System

Item # 14360-B

www.drivemedical.com

USER MANUAL

Item # 14360-P

Summary of Contents for 14360-B

Page 1: ...Med Aire Edge Alternating Pressure Low Air Loss Mattress Replacement System Item 14360 B www drivemedical com USER MANUAL Item 14360 P...

Page 2: ...l Unit 7 Mattress 8 PRODUCT FUNCTION 9 Control Unit 9 Mattress 10 OPERATION 12 General 12 CPR function 13 Pressure Range Selection 13 PATIENT POSITIONING AND COMFORT 14 CLEANING AND MAINTENANCE 15 TRO...

Page 3: ...The user is responsible for reading and understanding the product user manual Drive DeVilbiss Healthcare is not responsible for any injuries resulting from failure to comply with the instructions and...

Page 4: ...it from its power supply and allow at least 24 hours for drying Warning Drive DeVilbiss Healthcare support surfaces are designed as mattress replacement systems The risk of entrapment may occur when...

Page 5: ...needs recognizing that pressure prevention devices are only one component of a comprehensive pressure injury management program Support surfaces are not substitutes for turning repositioning or functi...

Page 6: ...with spinal cord injury once the acute injury has been stabilized and these patients have been assessed and cleared by the appropriate clinician Contraindications Patient conditions for which the appl...

Page 7: ...log control unit includes an easy to use pressure dial that is adjustable to the patient s weight and comfort Static button is available to discontinue alternation therapy for patient transfers caregi...

Page 8: ...tient is positioned properly within the confines of the bed The patient s head should be positioned in the center of the top section of the therapy mattress Do not let any extremities protrude over th...

Page 9: ...n on the panel the pump will start stop operation Cycle Time 2 The cycle time can be selected from the panel to choose the appropriate cycle time of the inflation modes The cycle time value options ar...

Page 10: ...n for 3 seconds to lock the panel press again for 3 seconds to unlock the panel Mattress The Med Aire Edge Mattress Replacement comes with a hose connection at the foot end of the mattress and 20 indi...

Page 11: ...are pulled tightly DO NOT SECURE TO THE SIDE RAILS STRAPS WILL TEAR OFF a Position the control unit by hanging hooks over foot board of the bed b Attach the air hoses to the therapy mattress securely...

Page 12: ...ignal The information signal LED will continue flashing Press the Mute button again to re enable the audible information signal Step 3 Press the static button to engage alternating therapy mode Adjust...

Page 13: ...to alert a low pressure condition Press the Mute button to mute the information signal The information signal LED will continue flashing Press the Mute button again to re enable the audible informatio...

Page 14: ...four 4 fingers stacked vertically beneath the air cell directly underneath the sacral region 3 Ensure that the 4 fingers can slide with minimal resistance between the patients sacral region and the lo...

Page 15: ...ith a clean cloth to remove any dust or particles Make sure all areas are clean top and bottom both sides 4 Spray cloth with cleaning solution and clean faceplate and control unit casing DO NOT allow...

Page 16: ...plug to see if there are abrasions or excessive wear Check the mattress cover for signs of wear or damage Ensure the mattress cover and tubes are connected correctly Plug in the control unit and chec...

Page 17: ...not work thus mattress does not alternate and alternate failure alarm is activated 1 Synchronous motor is out of order 2 Wires inside synchronous motor not connect well 3 Lower PCB is out of order 1...

Page 18: ...supply 120 60Hz Size 35 W x 80 L x 6 H Air output 8 liter min Top cover Nylon PU with zipper Pressure range 20 mmHg 55 mmHg Air cells Nylon PU Cycle time 10 15 20 25 min Base Durable Nylon PVC 840D C...

Page 19: ...If you have a question about your Drive device or this warranty please contact an authorized Drive dealer 14360 P 18 months for control unit and mattress Your Drive brand product is warranted to be fr...

Page 20: ...99 Seaview Boulevard Port Washington NY 11050 Phone 516 998 4600 Fax 516 998 4601 www drivemedical com REV 0 02 20...