P1752 Owner's Manual| REV 002
April 21, 2015
The information in this document is subject to change without notice.
Page 32
Replace tie-wraps to hold power cord to frame
Plug mains cable into wall outlet
Test function
WARNING
When routing cables in the medical bed, precautions shall be taken to avoid
squeezing those between parts of the medical bed.
TROUBLESHOOTING GUIDE
WARNING
Before doing any repairs or maintenance to the bed frame, read all instructions,
cautions, and warnings. Repairs should be done by a skilled technician.
Effect
Possible Cause
Verification
Corrective
Action
Bed does not
stay in place
Caster lock is not
engaged
Caster locks not in locked
position
Activate caster
Lock
There may be an object in-
between the floor and
casters, the floor may be
slippery
Clean the floor;
remove any
objects that
may be in the
way. Make sure
the floor is dry.
Caster lock is not
functioning
Caster lock stuck in one
position
Contact Primus
Medical by Drive
1-877-224-0946
Actuators not
working
Wire connections
may be loose or
damaged
Visually check wire
connections for looseness
or fraying
Reconnect any
loose wires
and/or power
cords. If cord is
damaged
replace
immediately