ANNUAL INSPECTION
Perform all Quarterly and Semi-Annual inspections plus:
Mattress Support Surface, Frame and Base Assemblies
Inspect welds on the mattress support surface, frame and base assemblies for
stress fractures.
Verify all fasteners are tight.
Inspect fasteners for wear or damage.
Inspect bed and Assist Rails bolts, if loose tighten and if missing replace.
Lubricate clevis pins at hinge points.
Lubricate Hi/Lo tracks of bed for smooth travel.
Actuators
Inspect push tubes and end connections of all actuators for excess wear or
bending.
SERVICING
DANGER SHOCK HAZARD
Possible Shock Hazard may occur if the main power supply cord for the Control Box
is not unplugged from the wall outlet before any maintenance is performed on
electrical components.
ACTUATORS AND CONTROL BOX
Cord and Socket Identification
Bed End Panel (Green)
Hand Pendant (Red)
Head Section Motor (Black)
Foot Section Motor (Yellow)
Foot End Hi-Lo Motor (Blue)
Head End Hi-Lo Motor (White)
Battery Back-Up (Black)