P1752 Owner's Manual| REV 002
April 21, 2015
The information in this document is subject to change without notice.
Page 28
Contact Primus Medical by Drive if you need any help with installation, usage
or maintenance of the product.
Corporate Office: Drive Medical, 99 Seaview Blvd.,
Port Washington, NY 11050. Ph: 1-877-224-0946 Fax: 516-988-4601
Website:
Manufacturer: Land America Health & Fitness Co., Ltd
Cleaning Instructions
If possible, remove resident before cleaning bed.
Prior to cleaning unplug power supply cords
Ensure all electrical parts (motors, control boxes, wires and pendant) are not
broken. Ensure that NO liquids enter electrical components.
Remove all gross/solid contaminants, then wash and sanitize all components.
DO NOT submerge the bed frame or electrical components.
Solvents, alcohol, bleach, caustic agents, high acid or alkaline solutions or
petroleum based products should not be used on the bed.
Use standard water pressure. DO NOT power wash or steam clean any parts.
Rinse completely with water (Maximum temperature 110
o
F or 43
o
C). DO NOT
rinse with cold water.
Ensure all parts are dry before using or storing.
NOTE:
This bed is NOT compatible with wash down tunnels.
INSPECTIONS
WARNING
Failure to properly maintain your bed may decrease the life expectancy of your
product, increase product maintenance requirements and costs and increase risk to
residents and staff.
Always service the bed at the required intervals.
INITIAL INSPECTION
Inspection of All Components – Receipt of assembled bed
Check bed components for obvious damage.
Inspect power supply cords for cuts and/or damage.
Check that actuator cords are connected properly to the controller.
Verify proper functionality of all features.
QUARTERLY INSPECTION
Check bed for proper function
Inspect all welds for cracks.
Inspect bed and Assist Rail bolts, if loose tighten and if missing replace.
Lubricate clevis pins at hinge points.
Lubricate tracks of bed for smooth travel.
Ensure cables routed in the medical bed, will not be squeezed between parts
of the medical bed.
Check all electrical cable strain reliefs to ensure there are no cracks or
breaks.